Harmonics Mode
- Display
- Bargraph showing 1st up to 33rd harmonic and DC; amplitude displayed in % relative to fundamental
- Analyzed waveform
- Voltage waveform(Ch.A), Current wavefrom (Ch.B) or Power (Ch.A x Ch.B), automatically generated
- Harmonics frequency range
- DC up to 33rd harmonic (for fundamental up to 60 Hz); DC up to 24th harmonic (for fundamental up to 400 Hz)
- Measurements
- Relative amplitude of individual harmonics; THD in %r or %f
Mechanical And General Specifications
- Warranty
- 3-years on main instrument 1-year on accessories
- Weight
- 1.2 kg
- Size
- 232 x 115 x 50 mm
- RS232/USB interface
- Optically isolated via optional PM9080 or OC4USB for interfacing to Printer or PC
- Battery life
- 7 hrs using NiMH
Bus Health Test Functions
- Bus systems supported
- AS-i (EN50295, 166 kb/s); CAN-bus (ISO-11898, up to 1 Mb/s); Interbus S (EIA-485, 500 kb/s); ControlNet (61158 type 2, 5Mb/s); Modbus (EIA-232 up to 115 kb/s and EIA-485 up to 10 Mb/s); Foundation Fieldbus H1 (61158 type 1, 31.25 kb/s); Profibus DP (EIA-485 up to 10 Mb/s); Profibus PA (61158 type 1, 31.25 kb/s); Ethernet 10Base2 (coaxial); Ethernet 10BaseT (UTP, 10 Mb/s); RS-232 (EIA-232, up to 115 kb/s); RS-485 (EIA-485, up to 10 Mb/s); User defined single wire or balanced system
- Eyepattern display mode
- Gives waveform display of bus signals with display persistence Timebase and attenuator settings may be altered by the user
- Measured parameters (where applicable)
- Baud rate, Risetime, Fall time, High level, Low level, Bias voltage level, Signal amplitude, Distortion, Jitter
- Parameter classification
- Default values: well within limits = 'good' within certain percentage of the limit values = 'weak' beyond limit values = 'bad' Limits values, as a default, are based on industry standard for the selected bus type or may be set by the user.
- Modes
- Waveform Parameter Analysis With Automatic Read-out And Parameter: validation (good / weak / bad); test limits are based on industry standards or may be set by the user Eyepattern display mode
- Bus health test
- Verifies the electrical parameters of industrial bus systems using automatic measurement and analysis functions. Next to that, an Eyepattern mode is provided for visual inspection of signal quality.
Specifications
- Input sensitivity
- 5 mV - 500 V/div
- TrendPlot recording
- Automatic vertical scaling and time compression Display the actual, MIN, MAX and AVG reading
- Maximum record length
- 512 min/max pts per input
- Memory
- 20 screens and setups
- Oscilloscope bandwidth
- 40 MHz
- Trigger types
- Connect-and-View™, Free Run, Single Shot, Edge, Video, External
- Scope measurements
- Automatic measurements: 26 (all instruments) On top, Fluke 125 has additional measurement functions for Power (W), VA, VAR and Power Factor (PF) for single phase and balanced-three-phase (delta) power systems
- Harmonics mode
- Capability to analyze the harmonics contents of power signals (see Harmonics Mode Section below for details)
- Dual input true RMS meter
- VDC, VAC, VAC + VDC, Ohms,Continuity, Diode-test Current, °C, °F, Capacitance, dBV, dBM, Crest Factor, Touch Hold and Zeroset
- Cursors
- Fluke 125/124: ∆T, 1/∆T, V, ∆V, rise- and fall time
- Timebase range
- 20 ns - 1 min/div
- Sample rate
- 2.5 GS/s repetitive sampling 25 MS/s single shot sampling
- Glitch capture
- 40 ns
- Inputs and digitizers
- 2
Safety Specifications
- Electrical safety
- 600V CAT III (EN61010-1)
Environmental Specifications
- Operating temperature
- 0°C to 50°C
- Storage temperature
- -20°C to +60°C
- Operating altitude
- 3,000 m
Additional Power Measurement Capabilities
- Current measurement
- Channel B, using i400s (included) or other compatible current clamp
- Measurement types
- Watt, VA VAR, Power Factor (PF)
- Current clamp or shunt sensitivity
- 0.1 / 1 / 10 / 100 / 1000 mV/A, 10 mV/mA and 400 mV/A
- Voltage measurement
- Channel A; using STL120, voltage probe or direct input
- Power configuration
- Single phase or balanced three-phase (delta-configuration) mains supply
Andere Kategorien
AktuatorenBlöckeKabel, Rohre & SchläucheKompressorenKontrolleKupplungen & FittingsZylinder & KolbenElektronikFilterHVACIndustrie-PCs & HMIsBeleuchtungLineartechnikMotorenStromversorgungLeistungsübertragungPumpen & VentileErneuerbare energieRobotikDichtungen, Ringe & VerbinderSensoren und MessgeräteWerkzeuge und ZubehörCNCSonstiges