Fluke

Fluke 9142-A-P-156

1 / 2
thumbnail
thumbnail
Part Number
Fluke 9142-A-P-156
  • Erhalten Sie bis zu 90 Tage Zahlungsziele von 500+ Lieferanten
  • Wiederverkaufsrabatte bis zu 40%!
Sign-Up
Preisanfrage senden
1
Zahlungsmöglichkeiten:
Fordern Sie ein Angebot an und wir senden Ihnen Informationen zu Verfügbarkeit und Preis.

P Specifications

Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1
±0.010°C at -25°C±0.015°C at 0°C±0.020°C at 50°C±0.025°C at 150°C±0.030°C at 200°C±0.040°C at 350°C±0.050°C at 420°C±0.070°C at 660°C
Reference resistance range
0 ohms to 400 ohms
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
Reference measurement capability
4-wire
Reference probe connection
6-pin Din with Infocon Technology
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°CPT-100 (385): ±0.02°C at 0°CPT-100 (3926): ±0.02°C at 0°CPT-100 (JIS): ±0.02°C at 0°C
RTD resistance range
0 ohms to 400 ohms
RTD resistance accuracy2
0 ohms to 25 ohms: ±0.002 ohms25 ohms to 400 ohms: ±80 ppm of reading
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
RTD connection
4 terminal input
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°CType K: ±0.8°C at 660°CType T: ±0.8°C at 400°CType E: ±0.7°C at 660°CType R: ±1.4°C at 660°CType S: ±1.5°C at 660°CType M: ±1.4°C at 660°CType L: ±0.7°C at 660°CType U: ±0.75°C at 600°CType N: ±0.9°C at 660°CType C: ±1.1°C at 660°C
TC millivolt range
–10 mV to 75 mV
Voltage accuracy
0.025% of reading + 0.01 mV
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
TC connection
Small connectors
Built-in mA readout accuracy
0.02% of reading + 2 mV
MA range
Cal 4-22 mA, Spec 4-24 mA
MA connection
2 terminal input
Loop power function
24 V DC loop power
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C

Base Unit Specifications

Temperature range at 23°C
–25°C to 150°C (–13°F to 302°F)
Display accuracy
±0.2°C full range
Stability
±0.01°C full range
Axial uniformity at 40 mm (1.6 in)
±0.05°C full range
Radial uniformity
±0.01°C full range
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C full range
Hysteresis
0.025
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Immersion (well) Depth
150 mm (5.9 in)
Insert OD
30 mm (1.18 in)
Heating time
16 min: 23°C to 140°C23 min: 23°C to 150°C25 min: –25°C to 150°C
Cooling time
15 min: 23°C to –25°C25 min: 150°C to –23°C
Resolution
0.01°
Display
LCD, °C or °F user-selectable
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Weight
8.16 kg (18 lb)
Power requirements
100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W
Computer interface
RS-232 interface
Calibration
NVLAP accredited calibration included
Automation
Use the 1586A Super-DAQ to automate temperature sensor calibration