P Specifications
- TC connection
- Small connectors
- Built-in TC thermometer readout accuracy
- Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
- RTD resistance accuracy2
- 25 ohms to 400 ohms: ±80 ppm of reading 0 ohms to 25 ohms: ±0.002 ohms
- Voltage accuracy
- 0.025% of reading + 0.01 mV
- Reference resistance accuracy2
- 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
- Built-in RTD thermometer readout accuracy
- NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
- RTD measurement capability
- 4-wire RTD (2-,3-wire RTD w/ Jumpers only)
- Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
- 1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
- Loop power function
- 24 V DC loop power
- RTD connection
- 4 terminal input
- TC millivolt range
- –10 mV to 75 mV
- MA connection
- 2 terminal input
- Built-in mA readout accuracy
- 0.02% of reading + 2 mV
- RTD resistance range
- 0 ohms to 400 ohms
- Reference resistance range
- 0 ohms to 400 ohms
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- Built-in reference thermometer readout accuracy (4-wire reference probe)1
- ±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
- Reference characterizations
- ITS-90, CVD, IEC-751, Resistance
- Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per°C
- RTD characterizations
- PT-100 (385),(JIS),(3926), NI-120, Resistance
- Reference measurement capability
- 4-wire
- Reference probe connection
- 6-pin Din with Infocon Technology
Base Unit Specifications
- Hysteresis
- 0.1
- Stability
- ±0.03°C at 50°C ±0.04°C at 420°C±0.05°C at 660°C
- Display accuracy
- ±0.35°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C
- Temperature range at 23°C
- 50°C to 660°C (122°F to 1220°F)
- Immersion (well) depth
- 150 mm (5.9 in)
- Weight
- 7.7 kg (17 lb)
- Display
- LCD, °C or °F user-selectable
- Operating conditions
- 0°C to 50°C, 0% to 90% RH (non-condensing)
- Environmental conditions (for all specifications except temperature range)
- 13°C to 33°C
- Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
- ±0.015°C at 50°C ±0.025°C at 420°C ±0.035°C at 660°C
- Power requirements
- 100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
- Insert OD
- 24.4 mm (0.96 in)
- Computer interface
- RS-232 and 9930 Interface-it control software included
- Size (H x W x D)
- 290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
- Radial uniformity
- ±0.02°C at 50°C ±0.05°C at 420°C±0.10°C at 660°C
- Resolution
- 0.01°
- Heating time
- 15 min: 50°C to 660°C
- Cooling time
- 25 min: 660°C to 100°C 35 min: 660°C to 50°C
- Axial uniformity at 40 mm (1.6 in)
- ±0.05°C at 50°C ±0.35°C at 420°C±0.5°C at 660°C
Inne kategorie
AktuatoryBlokiPrzewody, rury i wężeSprężarkiKontrolaŁączniki i złączkiCylindry i tłokiElektronikaFiltracjaHVACKomputery przemysłowe i interfejsy HMIOświetlenieTechnika liniowaSilnikiZasilaczPrzekazywanie mocyPompy i zaworyEnergia odnawialnaRobotykaUszczelki, pierścienie i złączkiCzujniki i urządzenia pomiaroweNarzędzia i akcesoriaCNCInne