Fluke
9144-F-P-156
The 9144 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.
- Termin płatności do 90 dni od 500+ dostawców.
- Rabaty na odsprzedaż do 40%!
Wyślij zapytanie o cenę
1
Metody płatności:
Additional Information
- Part Number
- 9144-F-P-156
- UPC
- 095969420525
Base Unit Specifications
- Temperature range at 23°C
- 50°C to 660°C (122°F to 1220°F)
- Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
- ±0.015°C at 50°C ±0.025°C at 420°C ±0.035°C at 660°C
- Computer interface
- RS-232 and 9930 Interface-it control software included
- Display accuracy
- ±0.35°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C
- Cooling time
- 25 min: 660°C to 100°C 35 min: 660°C to 50°C
- Operating conditions
- 0°C to 50°C, 0% to 90% RH (non-condensing)
- Axial uniformity at 40 mm (1.6 in)
- ±0.05°C at 50°C ±0.35°C at 420°C±0.5°C at 660°C
- Immersion (well) depth
- 150 mm (5.9 in)
- Display
- LCD, °C or °F user-selectable
- Resolution
- 0.01°
- Insert OD
- 24.4 mm (0.96 in)
- Heating time
- 15 min: 50°C to 660°C
- Power requirements
- 100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
- Size (H x W x D)
- 290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
- Weight
- 7.7 kg (17 lb)
- Hysteresis
- 0.1
- Environmental conditions (for all specifications except temperature range)
- 13°C to 33°C
- Stability
- ±0.03°C at 50°C ±0.04°C at 420°C±0.05°C at 660°C
- Radial uniformity
- ±0.02°C at 50°C ±0.05°C at 420°C±0.10°C at 660°C
P Specifications
- Voltage accuracy
- 0.025% of reading + 0.01 mV
- MA connection
- 2 terminal input
- Reference characterizations
- ITS-90, CVD, IEC-751, Resistance
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- Loop power function
- 24 V DC loop power
- TC millivolt range
- –10 mV to 75 mV
- Built-in mA readout accuracy
- 0.02% of reading + 2 mV
- RTD characterizations
- PT-100 (385),(JIS),(3926), NI-120, Resistance
- Reference resistance range
- 0 ohms to 400 ohms
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- TC connection
- Small connectors
- Built-in TC thermometer readout accuracy
- Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
- RTD connection
- 4 terminal input
- RTD measurement capability
- 4-wire RTD (2-,3-wire RTD w/ Jumpers only)
- Reference probe connection
- 6-pin Din with Infocon Technology
- RTD resistance accuracy2
- 25 ohms to 400 ohms: ±80 ppm of reading 0 ohms to 25 ohms: ±0.002 ohms
- Reference measurement capability
- 4-wire
- Reference resistance accuracy2
- 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
- Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per°C
- RTD resistance range
- 0 ohms to 400 ohms
- Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
- 1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
- Built-in RTD thermometer readout accuracy
- NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
- Built-in reference thermometer readout accuracy (4-wire reference probe)1
- ±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
Inne kategorie
AktuatoryBlokiPrzewody, rury i wężeSprężarkiKontrolaŁączniki i złączkiCylindry i tłokiElektronikaFiltracjaHVACKomputery przemysłowe i interfejsy HMIOświetlenieTechnika liniowaSilnikiZasilaczPrzekazywanie mocyPompy i zaworyEnergia odnawialnaRobotykaUszczelki, pierścienie i złączkiCzujniki i urządzenia pomiaroweNarzędzia i akcesoriaCNCInne