Fluke
9142-D-156
1 / 2
- Part Number
- 9142-D-156
- UPC
- 095969419543
- Consigue plazos de pago de hasta 90 días de 500+ proveedores.
- Descuentos de reventa de hasta el 40%!
Enviar solicitud de precio
1
Formas de pago:
P Specifications
- Built-in mA readout accuracy
- 0.02% of reading + 2 mV
- Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1
- ±0.010°C at -25°C ±0.015°C at 0°C±0.020°C at 50°C±0.025°C at 150°C±0.030°C at 200°C±0.040°C at 350°C±0.050°C at 420°C ±0.070°C at 660°C
- RTD measurement capability
- 4-wire RTD (2-,3-wire RTD w/ Jumpers only)
- MA connection
- 2 terminal input
- Reference measurement capability
- 4-wire
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per °C
- Reference probe connection
- 6-pin Din with Infocon Technology
- Reference characterizations
- ITS-90, CVD, IEC-751, Resistance
- TC connection
- Small connectors
- RTD characterizations
- PT-100 (385),(JIS),(3926), NI-120, Resistance
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- 1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
- RTD connection
- 4 terminal input
- Reference resistance accuracy2
- 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
- RTD resistance accuracy2
- 0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- Loop power function
- 24 V DC loop power
- Reference resistance range
- 0 ohms to 400 ohms
- Built-in TC thermometer readout accuracy
- Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
- TC millivolt range
- –10 mV to 75 mV
- Voltage accuracy
- 0.025% of reading + 0.01 mV
- Built-in RTD thermometer readout accuracy
- NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
- RTD resistance range
- 0 ohms to 400 ohms
Base Unit Specifications
- Computer interface
- RS-232 and 9930 Interface-it control software included
- Resolution
- 0.01°
- Hysteresis
- 0.025
- Power requirements
- 100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W
- Axial uniformity at 40 mm (1.6 in)
- ±0.05°C full range
- Stability
- ±0.01°C full range
- Environmental conditions (for all specifications except temperature range)
- 13°C to 33°C
- Weight
- 8.16 kg (18 lb)
- Immersion (well) Depth
- 150 mm (5.9 in)
- Cooling time
- 15 min: 23°C to –25°C 25 min: 150°C to –23°C
- Operating conditions
- 0°C to 50°C, 0% to 90% RH (non-condensing)
- Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
- ±0.006°C full range
- Insert OD
- 30 mm (1.18 in)
- Display
- LCD, °C or °F user-selectable
- Radial uniformity
- ±0.01°C full range
- Size (H x W x D)
- 290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
- Temperature range at 23°C
- –25°C to 150°C (–13°F to 302°F)
- Heating time
- 16 min: 23°C to 140°C 23 min: 23°C to 150°C 25 min: –25°C to 150°C
- Display accuracy
- ±0.2°C full range
Otras categorías
ActuadoresBloquesCables, tuberías y manguerasCompresoresControlAcoplamientos y AccesoriosCilindros y PistonesElectrónicaFiltrosHVACPCs industriales y HMIsIluminaciónTécnica linealMotoresFuente de alimentaciónTransmisión de energíaBombas y VálvulasEnergías renovablesRobóticaSellos, anillos y conectoresSensores y dispositivos de mediciónHerramientas y AccesoriosCNCOtros