Fluke
9143-D-156
The 9143 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.
- Consigue plazos de pago de hasta 90 días de 500+ proveedores.
- Descuentos de reventa de hasta el 40%!
Enviar solicitud de precio
1
Formas de pago:
P specifications
- RTD characterizations
- PT-100 (385),(JIS),(3926), NI-120, Resistance
- MA connection
- 2 terminal input
- Built-in TC thermometer readout accuracy
- Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per °C
- Built-in RTD thermometer readout accuracy
- NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
- RTD resistance accuracy‡
- 0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
- TC millivolt range
- –10 mV to 75 mV
- Reference measurement capability
- 4-wire
- Built-in reference thermometer readout accuracy (4-Wire reference probe)1
- ±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
- Reference probe connection
- 6-pin Din with Infocon Technology
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- Voltage accuracy
- 0.025% of reading + 0.01 mV
- Reference resistance range
- 0 ohms to 400 ohms
- RTD connection
- 4 terminal input
- Loop power function
- 24 V DC loop power
- Reference resistance accuracy2
- 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
- Reference characterizations
- ITS-90, CVD, IEC-751, Resistance
- RTD measurement capability
- 4-wire RTD (2-,3-wire RTD w\ Jumpers only)
- Built-in mA readout accuracy
- 0.02% of reading + 2 mV
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- 1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
- TC connection
- Small connectors
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- RTD resistance range
- 0 ohms to 400 ohms
Base Unit Specifications
- Resolution
- 0.01°
- Computer interface
- RS-232 and 9930 Interface-it control software included
- Display accuracy
- ±0.2°C Full Range
- Stability
- ±0.02°C at 33°C ±0.02°C at 200°C±0.03°C at 350°C
- Radial uniformity
- ±0.01°C at 33°C ±0.015°C at 200°C±0.02°C at 350°C
- Hysteresis
- 0.03
- Insert OD
- 25.3 mm (1.00 in)
- Weight
- 7.3 kg (16 lb)
- Heating time
- 5 min: 33°C to 350°C
- Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
- ±0.015°C Full Range
- Environmental conditions (for all specifications except temperature range)
- 13°C to 33°C
- Power requirements
- 100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
- Temperature range at 23°C
- 33°C to 350°C (91°F to 662°F)
- Size (H x W x D)
- 290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
- Cooling time
- 32 min: 350°C to 33°C 14 min: 350°C to 100°C
- Axial uniformity at 40 mm (1.6 in)
- ±0.04°C at 33°C ±0.1°C at 200°C±0.2°C at 350°C
- Display
- LCD, °C or °F user-selectable
- Immersion (well) depth
- 150 mm (5.9 in)
- Operating conditions
- 0°C to 50°C, 0% to 90% RH (non-condensing)
Otras categorías
ActuadoresBloquesCables, tuberías y manguerasCompresoresControlAcoplamientos y AccesoriosCilindros y PistonesElectrónicaFiltrosHVACPC industriales y HMIsIluminaciónTécnica linealMotoresFuente de alimentaciónTransmisión de energíaBombas y VálvulasEnergías renovablesRobóticaSellos, anillos y conectoresSensores y dispositivos de mediciónHerramientas y AccesoriosCNCOtros