Solicite un presupuesto y le enviaremos información sobre disponibilidad y precio.

Operating Fluid

Mineral oil
Shell spindle oil 22, our reference 55-655, viscosity 40 cs @ 20°C

Pressure Ranges

Oil operated
Ranges to 20,000 psi (1400 bar)

Accuracy

Accuracy1
±0.015% of reading (±0.008% optional)

Weight Increments

Minimum standard weight increments
Up to 500 psi (35 bar): 1 psi (0.1 bar)500 to 5000 psi (35 to 350 bar): 10 psi (1 bar)5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar)Model P3112: 1 psi (0.1 bar)
Optional fractional weights
Up to 500 psi (35 bar): 0.1 psi (0.01 bar)500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar)5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar)

Materials of Construction

Standard weight material
Series 3 non-magnetic austenitic stainless steel
Weight density
2.7 g/cm³
Optional fractional weights
Solution heat treated aluminum
Piston material
Tungsten carbide with nickel binderDensity- 15.0 g/cm³
Cylinder material
Hardened martensitic steelTungsten carbide (water above 500 psi, 35 bar)
Thermal coefficients of expansion
Oil piston/cylinder 16.5 ppm/°CWater (above 500 psi, 35 bar) 11 ppm/°C

Options

PressCal software
Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments.
Fluids
Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory.

General

Test port adaptors
1/8, 1/4, 3/8 and 1/2 NPT or BSP
Instrument weight
36 lbs (16 kg)
Instrument size (W x D x H)
17.5 x 12 x 8.5
Mass set weight (typical)
80 lbs (36 kg)
Reservoir volume
150 cc
Screw press displacement
5.5 cc
Pump displacement
4.7 cc per stroke
O-ring seal materials