Fluke
Fluke P3111-1
1 / 1
- Part Number
- Fluke P3111-1
- Consigue plazos de pago de hasta 90 días de 500+ proveedores.
- Descuentos de reventa de hasta el 40%!
Enviar solicitud de precio
1
Formas de pago:
Operating Fluid
- Mineral oil
- Shell spindle oil 22, our reference 55-655, viscosity 40 cs @ 20°C
Pressure Ranges
- Oil operated
- Ranges to 20,000 psi (1400 bar)
Accuracy
- Accuracy1
- ±0.015% of reading (±0.008% optional)
Weight Increments
- Minimum standard weight increments
- Up to 500 psi (35 bar): 1 psi (0.1 bar)500 to 5000 psi (35 to 350 bar): 10 psi (1 bar)5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar)Model P3112: 1 psi (0.1 bar)
- Optional fractional weights
- Up to 500 psi (35 bar): 0.1 psi (0.01 bar)500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar)5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar)
Materials of Construction
- Standard weight material
- Series 3 non-magnetic austenitic stainless steel
- Weight density
- 2.7 g/cm³
- Optional fractional weights
- Solution heat treated aluminum
- Piston material
- Tungsten carbide with nickel binderDensity- 15.0 g/cm³
- Cylinder material
- Hardened martensitic steelTungsten carbide (water above 500 psi, 35 bar)
- Thermal coefficients of expansion
- Oil piston/cylinder 16.5 ppm/°CWater (above 500 psi, 35 bar) 11 ppm/°C
Options
- PressCal software
- Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments.
- Fluids
- Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory.
General
- Test port adaptors
- 1/8, 1/4, 3/8 and 1/2 NPT or BSP
- Instrument weight
- 36 lbs (16 kg)
- Instrument size (W x D x H)
- 17.5 x 12 x 8.5
- Mass set weight (typical)
- 80 lbs (36 kg)
- Reservoir volume
- 150 cc
- Screw press displacement
- 5.5 cc
- Pump displacement
- 4.7 cc per stroke
- O-ring seal materials
- Buna N as standard, Viton and EPDM available
Otras categorías
ActuadoresBloquesCables, tuberías y manguerasCompresoresControlAcoplamientos y AccesoriosCilindros y PistonesElectrónicaFiltrosHVACPCs industriales y HMIsIluminaciónTécnica linealMotoresFuente de alimentaciónTransmisión de energíaBombas y VálvulasEnergías renovablesRobóticaSellos, anillos y conectoresSensores y dispositivos de mediciónHerramientas y AccesoriosCNCOtros