Operating Fluids

Operating fluids
Distilled or deionized water

Pressure Ranges

Distilled water operated
Ranges to 10,000 psi (700 bar)

Accuracy

Accuracy1
±0.015% of reading (±0.008% optional)
Accuracy1
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range.

Weight Increments

Minimum standard weight increments
Up to 500 psi (35 bar): 1 psi (0.1 bar) 500 to 5000 psi (35 to 350 bar): 10 psi (1 bar) 5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar) Model P3112: 1 psi (0.1 bar)
Optional fractional weights
Up to 500 psi (35 bar): 0.1 psi (0.01 bar) 500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar) 5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar)

Materials of Construction

Thermal coefficients of expansion
Oil piston/cylinder 16.5 ppm/°C Water (above 500 psi, 35 bar) 11 ppm/°C
Standard weight material
Series 3 non-magnetic austenitic stainless steel
Cylinder material
Hardened martensitic steel Tungsten carbide (water above 500 psi, 35 bar)
Weight density
2.7 g/cm³
Optional fractional weights
Solution heat treated aluminum
Weight density
7.8 g/cm³
Piston material
Tungsten carbide with nickel binder Density- 15.0 g/cm³

Options

PressCal software
Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments
Fluids
Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory.

General

Instrument size (W x D x H)
17.5 x 12 x 8.5 in (440 x 300 x 215 mm)
O-ring seal materials
Buna N as standard, Viton and EPDM available
Pump displacement
4.7 cc per stroke
Instrument weight
36 lbs (16 kg)
Mass set weight (typical)
80 lbs (36 kg)
Test port adaptors
1/8, 1/4, 3/8 and 1/2 NPT or BSP
Screw press displacement
5.5 cc
Reservoir volume
150 cc