P Specifications
- Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1
- ±0.010°C at -25°C±0.015°C at 0°C±0.020°C at 50°C±0.025°C at 150°C±0.030°C at 200°C±0.040°C at 350°C±0.050°C at 420°C±0.070°C at 660°C
- Reference resistance range
- 0 ohms to 400 ohms
- Reference resistance accuracy2
- 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
- Reference characterizations
- ITS-90, CVD, IEC-751, Resistance
- Reference measurement capability
- 4-wire
- Reference probe connection
- 6-pin Din with Infocon Technology
- Built-in RTD thermometer readout accuracy
- NI-120: ±0.015°C at 0°CPT-100 (385): ±0.02°C at 0°CPT-100 (3926): ±0.02°C at 0°CPT-100 (JIS): ±0.02°C at 0°C
- RTD resistance range
- 0 ohms to 400 ohms
- RTD resistance accuracy2
- 0 ohms to 25 ohms: ±0.002 ohms25 ohms to 400 ohms: ±80 ppm of reading
- RTD characterizations
- PT-100 (385),(JIS),(3926), NI-120, Resistance
- RTD measurement capability
- 4-wire RTD (2-,3-wire RTD w/ Jumpers only)
- RTD connection
- 4 terminal input
- Built-in TC thermometer readout accuracy
- Type J: ±0.7°C at 660°CType K: ±0.8°C at 660°CType T: ±0.8°C at 400°CType E: ±0.7°C at 660°CType R: ±1.4°C at 660°CType S: ±1.5°C at 660°CType M: ±1.4°C at 660°CType L: ±0.7°C at 660°CType U: ±0.75°C at 600°CType N: ±0.9°C at 660°CType C: ±1.1°C at 660°C
- TC millivolt range
- –10 mV to 75 mV
- Voltage accuracy
- 0.025% of reading + 0.01 mV
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- TC connection
- Small connectors
- Built-in mA readout accuracy
- 0.02% of reading + 2 mV
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- MA connection
- 2 terminal input
- Loop power function
- 24 V DC loop power
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per °C
Base Unit Specifications
- Temperature range at 23°C
- –25°C to 150°C (–13°F to 302°F)
- Display accuracy
- ±0.2°C full range
- Stability
- ±0.01°C full range
- Axial uniformity at 40 mm (1.6 in)
- ±0.05°C full range
- Radial uniformity
- ±0.01°C full range
- Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
- ±0.006°C full range
- Hysteresis
- 0.025
- Operating conditions
- 0°C to 50°C, 0% to 90% RH (non-condensing)
- Environmental conditions (for all specifications except temperature range)
- 13°C to 33°C
- Immersion (well) Depth
- 150 mm (5.9 in)
- Insert OD
- 30 mm (1.18 in)
- Heating time
- 16 min: 23°C to 140°C23 min: 23°C to 150°C25 min: –25°C to 150°C
- Cooling time
- 15 min: 23°C to –25°C25 min: 150°C to –23°C
- Resolution
- 0.01°
- Display
- LCD, °C or °F user-selectable
- Size (H x W x D)
- 290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
- Weight
- 8.16 kg (18 lb)
- Power requirements
- 100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W
- Computer interface
- RS-232 interface
- Calibration
- NVLAP accredited calibration included
- Automation
- Use the 1586A Super-DAQ to automate temperature sensor calibration
Altre categorie
AttuatoriBlocchiCavi, tubi e manichetteCompressoriControlloGiunzioni e raccordiCilindri e PistoniElettronicaFiltriHVACPC industriali e HMIsIlluminazioneTecnica lineareMotoriAlimentatoreTrasmissione di potenzaPompe e ValvoleEnergia rinnovabileRoboticaGuarnizioni, Anelli e ConnettoriSensori e dispositivi di misuraStrumenti e AccessoriCNCAltro