Fluke

P3116-1

1 / 1
thumbnail
Part Number
P3116-1
  • Ottenete termini di pagamento fino a 90 giorni da 500+ fornitori.
  • Sconti sulla rivendita fino al 40%!
Registrati
Invia una richiesta di prezzo
1
Metodi di pagamento:
Richiedete un preventivo e vi invieremo informazioni su disponibilità e prezzi.

Operating Fluid

Mineral oil
Shell spindle oil 22, our reference 55-655, viscosity 40 cs @ 20°C

Pressure Ranges

Oil operated
Ranges to 20,000 psi (1400 bar)

Accuracy

Accuracy1
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range.
Accuracy1
±0.015% of reading (±0.008% optional)

Weight Increments

Minimum standard weight increments
Up to 500 psi (35 bar): 1 psi (0.1 bar) 500 to 5000 psi (35 to 350 bar): 10 psi (1 bar) 5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar) Model P3112: 1 psi (0.1 bar)
Optional fractional weights
Up to 500 psi (35 bar): 0.1 psi (0.01 bar) 500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar) 5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar)

Materials of Construction

Optional fractional weights
Solution heat treated aluminum
Weight density
2.7 g/cm³
Piston material
Tungsten carbide with nickel binder Density- 15.0 g/cm³
Standard weight material
Series 3 non-magnetic austenitic stainless steel
Thermal coefficients of expansion
Oil piston/cylinder 16.5 ppm/°C Water (above 500 psi, 35 bar) 11 ppm/°C
Cylinder material
Hardened martensitic steel Tungsten carbide (water above 500 psi, 35 bar)
Weight density
7.8 g/cm³

Options

Fluids
Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory.
PressCal software
Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments.

General

Instrument weight
36 lbs (16 kg)
Reservoir volume
150 cc
O-ring seal materials
Buna N as standard, Viton and EPDM available
Screw press displacement
5.5 cc
Mass set weight (typical)
80 lbs (36 kg)
Pump displacement
4.7 cc per stroke
Instrument size (W x D x H)
17.5 x 12 x 8.5
Test port adaptors
1/8, 1/4, 3/8 and ½ NPT or BSP