Fluke

9143-D-P-156

1 / 2
thumbnail
thumbnail
Part Number
9143-D-P-156
UPC
095969419987
  • Ottenete termini di pagamento fino a 90 giorni da 500+ fornitori.
  • Sconti sulla rivendita fino al 40%!
Registrati
Invia una richiesta di prezzo
1
Metodi di pagamento:
Richiedete un preventivo e vi invieremo informazioni su disponibilità e prezzi.

P specifications

Reference probe connection
6-pin Din with Infocon Technology
RTD resistance accuracy‡
0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
Loop power function
24 V DC loop power
TC millivolt range
–10 mV to 75 mV
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Voltage accuracy
0.025% of reading + 0.01 mV
RTD measurement capability
4-wire RTD (2-,3-wire RTD w\ Jumpers only)
MA range
Cal 4-22 mA, Spec 4-24 mA
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
TC connection
Small connectors
Reference resistance range
0 ohms to 400 ohms
MA connection
2 terminal input
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
Built-in mA readout accuracy
0.02% of reading + 2 mV
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C
Built-in reference thermometer readout accuracy (4-Wire reference probe)1
±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
Reference measurement capability
4-wire
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
RTD resistance range
0 ohms to 400 ohms
RTD connection
4 terminal input

Base Unit Specifications

Cooling time
32 min: 350°C to 33°C 14 min: 350°C to 100°C
Computer interface
RS-232 and 9930 Interface-it control software included
Immersion (well) depth
150 mm (5.9 in)
Power requirements
100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
Radial uniformity
±0.01°C at 33°C ±0.015°C at 200°C±0.02°C at 350°C
Stability
±0.02°C at 33°C ±0.02°C at 200°C±0.03°C at 350°C
Display
LCD, °C or °F user-selectable
Heating time
5 min: 33°C to 350°C
Hysteresis
0.03
Display accuracy
±0.2°C Full Range
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.015°C Full Range
Insert OD
25.3 mm (1.00 in)
Axial uniformity at 40 mm (1.6 in)
±0.04°C at 33°C ±0.1°C at 200°C±0.2°C at 350°C
Temperature range at 23°C
33°C to 350°C (91°F to 662°F)
Resolution
0.01°
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Weight
7.3 kg (16 lb)
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)