Meter Readings

Maximum number of events
1024
Measurement Speed
Maximum 2 measurement/s
Recorded Time Span
2 weeks
Record Size (min, max, average)
2 M readings for 1 channel

Power (125B only)

VA reactive (var)
√((VA)2-W2)
VA
Vrms x Arms
Configurations
1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only, AUTOSET mode only)
Watt
RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes)
Full scale reading
999 counts
Full scale reading
999 counts
Power factor (PF)
Ratio between watts and VA range - 0.00 to 1.00
Full scale reading
999 counts

Crest factor (CREST)

Full scale reading
90 Counts
Range
1 to 10

Waveform Record

Recorded Time Span SD card
11 hours at 500 μs/div 14 days at 20 ms/div
Record Size SD card
1.5 G samples
Maximum sample rate
400 K sample/s
Size Internal memory
400 M samples Recorded Time
Span internal memory
15 minutes at 500 μs/div 11 hours at 20 ms/div
Maximum number of events
64

Decibel (dB)

0 dBm (600 Ω /50 Ω)
1 mW referenced to 600 Ω or 50 Ω
DB on
V dc, V ac, or Vac+dc
Full scale reading
1000 counts
0 dBV
1 V

Advanced Meter Functions

AutoHold (on A)
Captures and freezes a stable measurement result. Beeps when stable. AutoHold works on the main meter reading, with thresholds of 1 Vpp for AC signals and 100 mV for DC signals.
Zero Set
Set actual value to reference
Fixed decimal point
Activated by using attenuation keys

Pulse width (PULSE)

Frequency range
15 Hz (1 Hz) to 30 MHz in continuous autoset
Full scale reading
1000 counts

Input A and Input B

Accuracy
±(0.5% +5 counts)
Ranges
500 mV, 5 V, 50 V, 500 V, 750 V
Common mode rejection (CMRR)
>100 dB @ dc, >60 dB @ 50, 60, or 400 Hz
Full scale reading
5000 counts

Phase

Range
0 to 359 degrees
Resolution
1 degree
Modes
A to B, B to A

Continuity (Cont)

Measurement current
0.5 mA
Beep
<(30 Ω ±5 Ω) in 50 Ω range
Detection of shorts of
≥1 ms

Capacitance (CAP)

Ranges
50 nF, 500 nF, 5 μF, 50 μF, 500 μF
Measurement current
500 nA to 0.5 mA, increases with increasing ranges
Full scale reading
5000 counts

Frequency (Hz)

Full scale reading
10,000 counts
Frequency range
15 Hz (1 Hz) to 50 MHz in continuous autoset
Ranges
124B and 125B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz, 10 MHz, and 70 MHz
Accuracy @1 Hz to 1 MHz
±(0.5% +2 counts)
Ranges
123B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz,1 MHz, 10 MHz, and 50 MHz

Cursor Readout (124B and 125B)

Dual vertical lines
Peak-peak, time distance and reciprocal time distance readout
Single vertical line
Average, min and max readout
Dual horizontal lines
High, low and peak-peak readout
Rise or fall time
Transition time, 0%-level and 100%-level readout (manual or auto leveling; auto leveling only  possible in single channel mode)
Dual vertical lines
Average, min, max and time distance readout (in ROLL mode; instrument in HOLD)
Single vertical line
Average, min, max and time from start of readout (in ROLL mode; instrument in HOLD)
Sources
A, B
Single vertical line
Harmonics values in POWER QUALITY mode.
Single vertical line
Min, max and time from start of readout (in RECORDER mode; instrument in HOLD)
Accuracy
As oscilloscope accuracy

RPM

Accuracy
±(0.5% +2 counts)
Max reading
50.00 kRPM

Duty cycle (PULSE)

Frequency range
15 Hz (1 Hz) to 30 MHz in continuous autoset
Range
2% to 98%

Temperature (TEMP) with optional temperature probe

Accuracy
As V dc (add temp. probe uncertainty)
Range
200 °C/div (200 °F/div)
Scale factor
1 mV/°C and 1 mV/°F

Vpwm

Principle
Readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental frequency
Accuracy
As Vrms for sinewave signals
Purpose
To measure on pulse width modulated signals, like motor drive inverter outputs