Meter Readings

Measurement Speed
Maximum 2 measurement/s
Record Size (min, max, average)
2 M readings for 1 channel
Maximum number of events
1024
Recorded Time Span
2 weeks

Power (125B only)

Full scale reading
999 counts
Configurations
1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only, AUTOSET mode only)
Power factor (PF)
Ratio between watts and VA range - 0.00 to 1.00
VA
Vrms x Arms
Watt
RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes)
Full scale reading
999 counts
Full scale reading
999 counts
VA reactive (var)
√((VA)2-W2)

Crest factor (CREST)

Range
1 to 10
Full scale reading
90 Counts

Waveform Record

Span internal memory
15 minutes at 500 μs/div 11 hours at 20 ms/div
Recorded Time Span SD card
11 hours at 500 μs/div 14 days at 20 ms/div
Size Internal memory
400 M samples Recorded Time
Maximum sample rate
400 K sample/s
Record Size SD card
1.5 G samples
Maximum number of events
64

Decibel (dB)

0 dBV
1 V
Full scale reading
1000 counts
0 dBm (600 Ω /50 Ω)
1 mW referenced to 600 Ω or 50 Ω
DB on
V dc, V ac, or Vac+dc

Advanced Meter Functions

AutoHold (on A)
Captures and freezes a stable measurement result. Beeps when stable. AutoHold works on the main meter reading, with thresholds of 1 Vpp for AC signals and 100 mV for DC signals.
Fixed decimal point
Activated by using attenuation keys
Zero Set
Set actual value to reference

Pulse width (PULSE)

Full scale reading
1000 counts
Frequency range
15 Hz (1 Hz) to 30 MHz in continuous autoset

Input A and Input B

Accuracy
±(0.5% +5 counts)
Common mode rejection (CMRR)
>100 dB @ dc, >60 dB @ 50, 60, or 400 Hz
Full scale reading
5000 counts
Ranges
500 mV, 5 V, 50 V, 500 V, 750 V

Phase

Range
0 to 359 degrees
Modes
A to B, B to A
Resolution
1 degree

Continuity (Cont)

Beep
<(30 Ω ±5 Ω) in 50 Ω range
Measurement current
0.5 mA
Detection of shorts of
≥1 ms

Capacitance (CAP)

Measurement current
500 nA to 0.5 mA, increases with increasing ranges
Ranges
50 nF, 500 nF, 5 μF, 50 μF, 500 μF
Full scale reading
5000 counts

Frequency (Hz)

Frequency range
15 Hz (1 Hz) to 50 MHz in continuous autoset
Accuracy @1 Hz to 1 MHz
±(0.5% +2 counts)
Ranges
124B and 125B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz, 10 MHz, and 70 MHz
Ranges
123B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz,1 MHz, 10 MHz, and 50 MHz
Full scale reading
10,000 counts

Cursor Readout (124B and 125B)

Single vertical line
Harmonics values in POWER QUALITY mode.
Dual horizontal lines
High, low and peak-peak readout
Sources
A, B
Single vertical line
Average, min and max readout
Single vertical line
Min, max and time from start of readout (in RECORDER mode; instrument in HOLD)
Accuracy
As oscilloscope accuracy
Single vertical line
Average, min, max and time from start of readout (in ROLL mode; instrument in HOLD)
Dual vertical lines
Average, min, max and time distance readout (in ROLL mode; instrument in HOLD)
Dual vertical lines
Peak-peak, time distance and reciprocal time distance readout
Rise or fall time
Transition time, 0%-level and 100%-level readout (manual or auto leveling; auto leveling only  possible in single channel mode)

RPM

Accuracy
±(0.5% +2 counts)
Max reading
50.00 kRPM

Duty cycle (PULSE)

Frequency range
15 Hz (1 Hz) to 30 MHz in continuous autoset
Range
2% to 98%

Temperature (TEMP) with optional temperature probe

Range
200 °C/div (200 °F/div)
Scale factor
1 mV/°C and 1 mV/°F
Accuracy
As V dc (add temp. probe uncertainty)

Vpwm

Accuracy
As Vrms for sinewave signals
Principle
Readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental frequency
Purpose