Fluke

1551A-20-DL

Accurate and repeatable to ±0.05°C over its full range, the 1551A “Stik” Thermometer is the “gold standard” of industrial temperature calibration.
1551A-20-DL
1551A-20-DL
1551A-20-DL
1551A-20-DL
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
€1,258.73
1
Payment methods:
A total of 3 Parts In-Stock from 1 Distributor

Specifications

Operating temperature range readout
-10°C to 50°C (14°F to 122°F)
Sensor type
100 Ω thin-film RTD
AC power
None
Damping
Moving average of the most recent 2 to 10 readings (ON/OFF, 2, 5, 10)
EMC compliance
EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class B
Sample rate
Selectable: 0.5 seconds, 1 second, 2 seconds (default 1 second)
Sample rate
Selectable: 0.5 seconds, 1 second, 2 seconds (default 1 second)
Optional data logging1
Up to 10,000 time-stamped readings stored to internal memory
Probe sheath material
Stainless steel
Probe hysteresis
±0.01°C
Response time
4.8 mm diameter probe (3/16 in): 14 seconds 6.35 mm diameter probe (1/4 in): 21 seconds
Temperature resolution
Selectable: 0.1, 0.01, 0.001 (default 0.01)
DC power
3-AAA Batteries, typical battery life of 300 hours without LCD backlight
Operating temperature range readout
-10°C to 50°C (14°F to 122°F)
Accuracy (1 year)
±0.05°C (±0.09°F)
Temperature range
-50°C to 160°C (-58°F to 320°F)
Temperature resolution
Selectable: 0.1, 0.01, 0.001 (default 0.01)
Characterization
1. See ordering information for optional data logging configurations
Calibration (included)
NVLAP-accredited, NIST-traceable
Electronics dimensions (H x W x D)
114 x 57 x 25 mm (4.5 x 2.25 x 1.0 in)
Characterization
CVD
Storage temperature
-20°C to 60°C (-4°F to 140°F) 0% to 95% RH, non-condensing
Weight
196 g (6.9 oz)
Logging intervals1
2 seconds, 5 seconds, 10 seconds, 30 seconds, or 60 seconds; 2 minutes, 5 minutes, 10 minutes, 30 minutes, or 60 minutes
Display units
°C, °F
Communications
RS-232 stereo jack (access calibration parameters only)
Enclosure rating
IP50
Probe temperature coefficient
0.00385 Ω/Ω/°C nominal
Sensor length
≤ 10 mm
Sensor position (from sheath tip)
3 mm (0.1 in)
Minimum immersion depth
7 cm (2.8 in)