Mechanical and General Specifications

Size
270 x 190 x 70 mm
Weight
2.2 kg
Warranty
3-years on main instrument, 1-year on accessories
Li-Ion battery
BP290 2400 mAh battery up to 4 hours included standard

DMM Specification

Memory
Screen + Setup: 30
Memory
> Recording: 10
Maximum resolution
5000 counts
Measurements
VDC, VAC rms, VAC + DC rms, Resistance, Continuity, Diode Test, Current (through clamp or shunt), Temperature °C or °F (using optional converter)
Meter inputs (also used as External Trigger Input for the Oscilloscope Mode)
4 mm banana inputs, fully isolated from scope inputs and from scope ground
Display
153 mm full color LCD with fast update rate

Included Accessories

Two channel models
> Handstrap (affixed to instrument)
Two channel models
Hangstrap
Two channel models
USB Interface Cable
Two channel models
BC190 Battery Charger/mains Adapter
Two channel models
VPS410 Voltage Probe Sets (one red, one blue)
Two channel models
BP290 2400 mAh Li-ion Battery
Warranty
Three years on main instrument, one year on accessories
Two channel models
TL175 TwistGuard™ safety designed Test Lead Set (red and black)
Two channel models
FlukeView ScopeMeter® demo package (restricted functionality)
Two channel models
Multi-Language User Manuals on CD-Rom

Specifications

Persistence
Digital persistence giving analog oscilloscope-like waveform decay
Maximum record length
ScopeRecord mode: 30,000 pts
Glitch capture
8 ns
Maximum record length
Scope mode: 10,000 pts per input
Number of channels
100 MHz: 2 channels
Real-time sample rate
1.25 GS/s
Timebase range
5 ns - 2 min/div
Bandwidth
100 MHz
Scope measurements
Automatic: 30
Input sensitivity
2 mV-100 V/div
Independently floating isolated inputs
Up to 1000 V CAT III/600 V CAT IV between inputs, references and ground
Waveform compare
Waveform reference with automatic Pass/Fail testing
Scope measurements
Cursor: 7
Trigger source
Any scope input channel
Trigger types
Connect-and-View™, Free Run, Single Shot, Edge, Delay, Dual Slope, Video, Video Line, Selectable pulse width, N-Cycle
Inputs
2 scope plus DMM input

Safety Specifications

Electrical safety
1,000 V CAT III/600 V CAT IV (EN61010-1)

Environmental Specifications

Operating temperature
0°C to 50°C Excluding Batteries
Operating altitude
Up to 2,000 m (6666 ft) for CAT IV 600 V, CAT III 1000 V
Operating altitude
Up to 3,000 m (10,000 ft) for CAT III 600 V, CAT II 1000 V
Operating temperature
0°C to 40°C Including Batteries
Storage temperature
-20°C to +60°C