9142-A-156

The 9142 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.
9142-A-156
9142-A-156
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

P Specifications

Reference resistance range
0 ohms to 400 ohms
RTD resistance range
0 ohms to 400 ohms
Loop power function
24 V DC loop power
RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C
MA range
Cal 4-22 mA, Spec 4-24 mA
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
TC millivolt range
–10 mV to 75 mV
MA connection
2 terminal input
RTD resistance accuracy2
0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
Reference probe connection
6-pin Din with Infocon Technology
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Built-in mA readout accuracy
0.02% of reading + 2 mV
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1
±0.010°C at -25°C ±0.015°C at 0°C±0.020°C at 50°C±0.025°C at 150°C±0.030°C at 200°C±0.040°C at 350°C±0.050°C at 420°C ±0.070°C at 660°C
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
TC connection
Small connectors
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
Voltage accuracy
0.025% of reading + 0.01 mV
Reference measurement capability
4-wire
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
RTD connection
4 terminal input

Base Unit Specifications

Radial uniformity
±0.01°C full range
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Cooling time
15 min: 23°C to –25°C 25 min: 150°C to –23°C
Display
LCD, °C or °F user-selectable
Hysteresis
0.025
Computer interface
RS-232 and 9930 Interface-it control software included
Power requirements
100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W
Temperature range at 23°C
–25°C to 150°C (–13°F to 302°F)
Weight
8.16 kg (18 lb)
Display accuracy
±0.2°C full range
Insert OD
30 mm (1.18 in)
Resolution
0.01°
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C full range
Stability
±0.01°C full range
Immersion (well) Depth
150 mm (5.9 in)
Axial uniformity at 40 mm (1.6 in)
±0.05°C full range
Heating time
16 min: 23°C to 140°C 23 min: 23°C to 150°C 25 min: –25°C to 150°C
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)