9142-D-P-156

The 9142 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.
9142-D-P-156
9142-D-P-156
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

Base Unit Specifications

Heating time
16 min: 23°C to 140°C 23 min: 23°C to 150°C 25 min: –25°C to 150°C
Weight
8.16 kg (18 lb)
Display
LCD, °C or °F user-selectable
Immersion (well) Depth
150 mm (5.9 in)
Axial uniformity at 40 mm (1.6 in)
±0.05°C full range
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C full range
Temperature range at 23°C
–25°C to 150°C (–13°F to 302°F)
Cooling time
15 min: 23°C to –25°C 25 min: 150°C to –23°C
Resolution
0.01°
Computer interface
RS-232 and 9930 Interface-it control software included
Power requirements
100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W
Insert OD
30 mm (1.18 in)
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Hysteresis
0.025
Stability
±0.01°C full range
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Display accuracy
±0.2°C full range
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Radial uniformity
±0.01°C full range

P Specifications

RTD resistance accuracy2
0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1
±0.010°C at -25°C ±0.015°C at 0°C±0.020°C at 50°C±0.025°C at 150°C±0.030°C at 200°C±0.040°C at 350°C±0.050°C at 420°C ±0.070°C at 660°C
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C
RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
RTD connection
4 terminal input
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
Loop power function
24 V DC loop power
TC connection
Small connectors
Voltage accuracy
0.025% of reading + 0.01 mV
MA range
Cal 4-22 mA, Spec 4-24 mA
Reference probe connection
6-pin Din with Infocon Technology
Reference measurement capability
4-wire
Reference resistance range
0 ohms to 400 ohms
RTD resistance range
0 ohms to 400 ohms
TC millivolt range
–10 mV to 75 mV
MA connection
2 terminal input
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
Built-in mA readout accuracy
0.02% of reading + 2 mV
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.