The 9142 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.
9142-F-P-156
9142-F-P-156
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

P Specifications

Loop power function
24 V DC loop power
Reference probe connection
6-pin Din with Infocon Technology
TC connection
Small connectors
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C
Reference resistance range
0 ohms to 400 ohms
RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
Voltage accuracy
0.025% of reading + 0.01 mV
TC millivolt range
–10 mV to 75 mV
RTD connection
4 terminal input
Reference measurement capability
4-wire
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
RTD resistance accuracy2
0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
MA range
Cal 4-22 mA, Spec 4-24 mA
Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1
±0.010°C at -25°C ±0.015°C at 0°C±0.020°C at 50°C±0.025°C at 150°C±0.030°C at 200°C±0.040°C at 350°C±0.050°C at 420°C ±0.070°C at 660°C
Built-in mA readout accuracy
0.02% of reading + 2 mV
RTD resistance range
0 ohms to 400 ohms
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
MA connection
2 terminal input
Reference characterizations
ITS-90, CVD, IEC-751, Resistance

Base Unit Specifications

Weight
8.16 kg (18 lb)
Power requirements
100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C full range
Cooling time
15 min: 23°C to –25°C 25 min: 150°C to –23°C
Axial uniformity at 40 mm (1.6 in)
±0.05°C full range
Hysteresis
0.025
Stability
±0.01°C full range
Display
LCD, °C or °F user-selectable
Radial uniformity
±0.01°C full range
Heating time
16 min: 23°C to 140°C 23 min: 23°C to 150°C 25 min: –25°C to 150°C
Display accuracy
±0.2°C full range
Insert OD
30 mm (1.18 in)
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Temperature range at 23°C
–25°C to 150°C (–13°F to 302°F)
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Computer interface
RS-232 and 9930 Interface-it control software included
Resolution
0.01°
Immersion (well) Depth
150 mm (5.9 in)
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)