P specifications

Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
Reference probe connection
6-pin Din with Infocon Technology
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
MA range
Cal 4-22 mA, Spec 4-24 mA
TC millivolt range
–10 mV to 75 mV
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
RTD resistance accuracy‡
0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
RTD measurement capability
4-wire RTD (2-,3-wire RTD w\ Jumpers only)
Loop power function
24 V DC loop power
Reference measurement capability
4-wire
RTD resistance range
0 ohms to 400 ohms
RTD connection
4 terminal input
Reference resistance range
0 ohms to 400 ohms
TC connection
Small connectors
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Voltage accuracy
0.025% of reading + 0.01 mV
Built-in reference thermometer readout accuracy (4-Wire reference probe)1
±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
MA connection
2 terminal input
Built-in mA readout accuracy
0.02% of reading + 2 mV
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C

Base Unit Specifications

Stability
±0.02°C at 33°C ±0.02°C at 200°C±0.03°C at 350°C
Immersion (well) depth
150 mm (5.9 in)
Insert OD
25.3 mm (1.00 in)
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.015°C Full Range
Resolution
0.01°
Power requirements
100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
Axial uniformity at 40 mm (1.6 in)
±0.04°C at 33°C ±0.1°C at 200°C±0.2°C at 350°C
Hysteresis
0.03
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Display
LCD, °C or °F user-selectable
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Cooling time
32 min: 350°C to 33°C 14 min: 350°C to 100°C
Temperature range at 23°C
33°C to 350°C (91°F to 662°F)
Heating time
5 min: 33°C to 350°C
Weight
7.3 kg (16 lb)
Computer interface
RS-232 and 9930 Interface-it control software included
Radial uniformity
±0.01°C at 33°C ±0.015°C at 200°C±0.02°C at 350°C
Display accuracy
±0.2°C Full Range