Fluke

9144-D-156

The 9144 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.
9144-D-156
9144-D-156
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

Base Unit Specifications

Resolution
0.01°
Insert OD
24.4 mm (0.96 in)
Temperature range at 23°C
50°C to 660°C (122°F to 1220°F)
Hysteresis
0.1
Display
LCD, °C or °F user-selectable
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.015°C at 50°C ±0.025°C at 420°C ±0.035°C at 660°C
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Display accuracy
±0.35°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C
Power requirements
100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
Weight
7.7 kg (17 lb)
Radial uniformity
±0.02°C at 50°C ±0.05°C at 420°C±0.10°C at 660°C
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Heating time
15 min: 50°C to 660°C
Cooling time
25 min: 660°C to 100°C 35 min: 660°C to 50°C
Stability
±0.03°C at 50°C ±0.04°C at 420°C±0.05°C at 660°C
Axial uniformity at 40 mm (1.6 in)
±0.05°C at 50°C ±0.35°C at 420°C±0.5°C at 660°C
Computer interface
RS-232 and 9930 Interface-it control software included
Immersion (well) depth
150 mm (5.9 in)
Environmental conditions (for all specifications except temperature range)
13°C to 33°C

P Specifications

Built-in mA readout accuracy
0.02% of reading + 2 mV
MA connection
2 terminal input
RTD resistance accuracy2
25 ohms to 400 ohms: ±80 ppm of reading 0 ohms to 25 ohms: ±0.002 ohms
Reference probe connection
6-pin Din with Infocon Technology
RTD connection
4 terminal input
RTD resistance range
0 ohms to 400 ohms
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per°C
TC millivolt range
–10 mV to 75 mV
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Loop power function
24 V DC loop power
Reference measurement capability
4-wire
Built-in reference thermometer readout accuracy (4-wire reference probe)1
±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
Reference resistance range
0 ohms to 400 ohms
Voltage accuracy
0.025% of reading + 0.01 mV
MA range
Cal 4-22 mA, Spec 4-24 mA
TC connection
Small connectors
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)