9144-E-156

The 9144 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.
9144-E-156
9144-E-156
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

P Specifications

TC millivolt range
–10 mV to 75 mV
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
Loop power function
24 V DC loop power
MA range
Cal 4-22 mA, Spec 4-24 mA
Built-in reference thermometer readout accuracy (4-wire reference probe)1
±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per°C
Reference resistance range
0 ohms to 400 ohms
TC connection
Small connectors
Reference probe connection
6-pin Din with Infocon Technology
Reference measurement capability
4-wire
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Built-in mA readout accuracy
0.02% of reading + 2 mV
RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
RTD resistance range
0 ohms to 400 ohms
MA connection
2 terminal input
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
RTD connection
4 terminal input
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Voltage accuracy
0.025% of reading + 0.01 mV
RTD resistance accuracy2
25 ohms to 400 ohms: ±80 ppm of reading 0 ohms to 25 ohms: ±0.002 ohms
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance

Base Unit Specifications

Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Radial uniformity
±0.02°C at 50°C ±0.05°C at 420°C±0.10°C at 660°C
Display
LCD, °C or °F user-selectable
Hysteresis
0.1
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Axial uniformity at 40 mm (1.6 in)
±0.05°C at 50°C ±0.35°C at 420°C±0.5°C at 660°C
Immersion (well) depth
150 mm (5.9 in)
Resolution
0.01°
Computer interface
RS-232 and 9930 Interface-it control software included
Temperature range at 23°C
50°C to 660°C (122°F to 1220°F)
Insert OD
24.4 mm (0.96 in)
Cooling time
25 min: 660°C to 100°C 35 min: 660°C to 50°C
Power requirements
100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
Weight
7.7 kg (17 lb)
Stability
±0.03°C at 50°C ±0.04°C at 420°C±0.05°C at 660°C
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.015°C at 50°C ±0.025°C at 420°C ±0.035°C at 660°C
Display accuracy
±0.35°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C
Heating time
15 min: 50°C to 660°C
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)