Fluke

9144-F-156

The 9144 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.
9144-F-156
9144-F-156
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

P Specifications

RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
RTD connection
4 terminal input
MA range
Cal 4-22 mA, Spec 4-24 mA
MA connection
2 terminal input
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Reference probe connection
6-pin Din with Infocon Technology
TC connection
Small connectors
Voltage accuracy
0.025% of reading + 0.01 mV
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per°C
Reference resistance range
0 ohms to 400 ohms
Built-in reference thermometer readout accuracy (4-wire reference probe)1
±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
Built-in mA readout accuracy
0.02% of reading + 2 mV
Reference measurement capability
4-wire
RTD resistance range
0 ohms to 400 ohms
RTD resistance accuracy2
25 ohms to 400 ohms: ±80 ppm of reading 0 ohms to 25 ohms: ±0.002 ohms
TC millivolt range
–10 mV to 75 mV
Loop power function
24 V DC loop power

Base Unit Specifications

Computer interface
RS-232 and 9930 Interface-it control software included
Heating time
15 min: 50°C to 660°C
Display
LCD, °C or °F user-selectable
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.015°C at 50°C ±0.025°C at 420°C ±0.035°C at 660°C
Stability
±0.03°C at 50°C ±0.04°C at 420°C±0.05°C at 660°C
Resolution
0.01°
Immersion (well) depth
150 mm (5.9 in)
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Weight
7.7 kg (17 lb)
Display accuracy
±0.35°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C
Temperature range at 23°C
50°C to 660°C (122°F to 1220°F)
Hysteresis
0.1
Insert OD
24.4 mm (0.96 in)
Power requirements
100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Cooling time
25 min: 660°C to 100°C 35 min: 660°C to 50°C
Radial uniformity
±0.02°C at 50°C ±0.05°C at 420°C±0.10°C at 660°C
Axial uniformity at 40 mm (1.6 in)
±0.05°C at 50°C ±0.35°C at 420°C±0.5°C at 660°C