Fluke

FLK-1587/MDT FC Kit

FLK-1587/MDT FC Kit
FLK-1587/MDT FC Kit
FLK-1587/MDT FC Kit
FLK-1587/MDT FC Kit
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

Capacitance

Range  Resolution
Accuracy ±(% of Rdg+Counts)
1000 nF  1 nF
±(1.2% + 2)
10.00 μF  0.01 μF
±(1.2% + 2)
100.0 μF  0.1 μF
±(1.2% ±90 counts)
9999 μF  1 μF
±(1.2% ±90 counts)

Frequency Measurement

Range Resolution
Accuracy ±(% of Rdg+Counts)
99.99 Hz
0.01 Hz : ±(0.1% + 1)
999.9 Hz
0.1 Hz : ±(0.1% + 1)
9.999 kHz
0.001 kHz : ±(0.1% + 1)
99.99 kHz
0.01 kHz : ±(0.1% + 1)

DC Millivolts Measurement

Range Resolution
Accuracy ±(% of Rdg + Counts)
600.0 mVdc
0.1 mV: ±(0.1% + 1)

Insulation Specifications

Measurement range
0.01 MΩ to 2 GΩ
Test voltages
50, 100, 250, 500, 1000 V
Test voltage accuracy
+20%, -0%
Short-circuit test current
1 mA nominal
Auto discharge
Discharge time <0.5 second for C = 1 μF or less
Live circuit detection
Inhibit test if terminal voltage > 30 V prior to initialization of test
Maximum capacitive load
Operable with up to 1 μF load

Mechanical and General Specifications

Size
124 x 61 x 27 mm
Warranty
1 year
Weight
200 g
Maximum conductor diameter
32 mm (1.25 in)
Maximum conductor size (busbar)
750 MCM
Power supply
From unit under test
Warranty
2 years
Output cable length
1.5 m (59 in)
Shrouded banana plugs
Yes

Continuity Test

Continuity indication
Continuous audible tone for testresistance below 25 Ω and off above 100 Ω. Maximum Reading; 1000 Ω
Open circuit voltage
<8.0 V
Short circuit current
1.0 mA typical
Overload protection
1000 V rms
Response time
>1 m sec

AC 45 Hz to 1000 Hz

Range
400 mA || 60 mA
Resolution
0.1 mA || 0.01 mA
Accuracy ±(% of Rdg+Counts)
±(1.5% + 2)¹ || ±(1.5% + 2)¹
Burden voltage (Typical)
2 mV/mA || 2 mV/mA

Electromagnetic Compatibility

International
IEC 61326-1:Portable Electromagnetic Environment; IEC 61326-2-2 CISPR 11: Group 1, Class A || Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself. || Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes.  There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances. || Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object. The equipment may not meet the immunity requirements of this standard when test leads and/or test probes are connected.
Korea (KCC)
Class A Equipment (Industrial Broadcasting & Communication Equipment) || Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC)
47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Enclosure protection
IEC 60529: IP40 (non-operating)

Diode Test

Diode test indication
Display voltage drop: 0.6 V at1.0 mA nominal test current:
Accuracy
±(2% + 3)

DC Voltage Measurement

Range Resolution
Accuracy ±(% of Rdg + Counts)
6.000 V dc
0.001 V: ±(0.09% + 2)
60.00 V dc
0.01 V: ±(0.09% + 2)
600.0 V dc
0.1 V: ±(0.09% + 2)
1000 V dc
1 V: ±(0.09% + 2)
Input impedance
10 MΩ (nominal), <100 pF
Normal mode rejection ratio
>60 dB @ 50 Hz or 60 Hz
Common mode rejection ratio
>120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance)

Ohms Measurement

Range Resolution
Accuracy +(% of Rdg+Counts)¹
600.0 Ω
0.1 Ω: ±(0.9% + 2)
6.000 kΩ
0.001 kΩ: ±(0.9% + 2)
60.00 kΩ
0.01 kΩ: ±(0.9% + 2)
600.0 kΩ
0.1 KΩ: ±(0.9% + 2)
6.000 MΩ
0.001 MΩ: ±(0.9% + 2)
50.0 MΩ ²
0.01 MΩ: ±(1.5% + 3)
Overload protection
1000 V rms or dc
Open circuit test voltage
<8.0 V dc
Short circuit current
<1.1 mA

DC

Range
400 mA || 60 mA
Resolution
0.1 mA || 0.01 mA
Accuracy ±(% of Rdg+Counts)
±(0.2% + 2) || ±(0.2% + 2)
Burden voltage (Typical)
2 mV/mA || 2 mV/mA
Overload
600 mA for 2 minutes maximum
Fuse protection for mA input
0.44 mA, 1000 V, IR 10 kA
AC conversion
Inputs are ac-coupled and calibrated to the rms value of sine wave input

Technical Specifications

Voltage range
40-700 V
Frequency range
15-400 Hz
Operating timer
Continuous

Temperature Measurement

Range  Resolution
Accuracy ±(% of Rdg+Counts)¹
-40 ° C to 537 ° C
0.1 °C : ±(1% + 10 counts)
-40 ° F to 998 ° F
0.1 °F : ±(1% + 18 counts)

Specifications

Nominal current range
400  A
Continuous current range
1 A - 400 A
Maximum non-destructive current
1000 A
Lowest measurable current
1 A
Basic accuracy
(% reading + floorspec)
Useable frequency
5 Hz - 20 kHz
Output level(s)
1 mA/A

Safety Specifications

Electrical safety
EN 61010,  EN 61557-7
Safety
CAT IV 600 V, CAT III, 1000 V
Overvoltage category
CAT IV, 300V
Maximum voltage
1000 V AC rms from earth to ground, in compliance with EN61010

Frequency Counter Sensitivity

600.0 mV ac
V ac Sensitivity (RMS Sine Wave)¹  5 Hz to 20 kHz : 100.0 mV  || V ac Sensitivity (RMS Sine Wave)¹  20 kHz to 100 kHz: 150.0 mV  || DC Trigger Levels¹ to 20 kHz²: NA
6.0 V
V ac Sensitivity (RMS Sine Wave)¹  5 Hz to 20 kHz : 1.0 V  || V ac Sensitivity (RMS Sine Wave)¹  20 kHz to 100 kHz: 1.5 V  || DC Trigger Levels¹ to 20 kHz²: -400.0 mV and 2.5 V
60.0 V
V ac Sensitivity (RMS Sine Wave)¹  5 Hz to 20 kHz : 10.0 V  || V ac Sensitivity (RMS Sine Wave)¹  20 kHz to 100 kHz: 36.0 V  || DC Trigger Levels¹ to 20 kHz²: 1.2 V and 4.0 V
600.0 V
V ac Sensitivity (RMS Sine Wave)¹  5 Hz to 20 kHz : 100.0 V  || V ac Sensitivity (RMS Sine Wave)¹  20 kHz to 100 kHz: - || DC Trigger Levels¹ to 20 kHz²: 12.0 V and 40.0 V
1000.0 V
V ac Sensitivity (RMS Sine Wave)¹  5 Hz to 20 kHz : 300.0 V  || V ac Sensitivity (RMS Sine Wave)¹  20 kHz to 100 kHz: - || DC Trigger Levels¹ to 20 kHz²: 12.0 V and 40.0 V

Environmental Specifications

Operating temperature
0 °C to 40 °C
Dust/water resistance
IP 40

General Specifications

Maximum voltage applied to any terminal and common
1000 V
Storage temperature
-40 °C to 60 °C (-40 °F to 140 °F)
Operating temperature
-20 °C to 55 °C (-4 °F to 131 °F)
Temperature coefficient
0.05 x (specified accuracy) per °C for temperatures <18 °C or >28 °C (<64 °F or >82 °F)
Relative humidity
Noncondensing || 0% to 95% @ 10 °C to 30 °C (50 °F to 86 °F) || 0% to 75% @ 30 °C to 40 °C (86 °F to 104 °F) || 0% to 40% @ 40 °C to 55 °C (104 °F to 131 °F)
Vibration
Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
Radio frequency communication
2.4 GHz ISM Band
Radio frequency certification
FCC: T68-FBLE, IC: 6627A-FBLE

Safety

IEC 61010-1
Pollution Degree 2
IEC 61010-2-033
CAT IV 600 V / CAT III 1000 V
Batteries
Four AA batteries (NEDA 15A or IEC LR6)
Battery life
Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off.
Size (H x W x L)
5.0 x 10.0 x 20.3 cm (1.97 x 3.94 x 8.00 in)
Weight
550 g (1.2 lb)
Altitude
Operating: 2000 m
Altitude Storage
12,000 m
Over-range capability
110% of range except for capacitance which is 100%
Frequency overload protection
<107 V Hz
Fuse protection for mA input
0.44A, 1000 V, IR 10 kA

Range

600.0 mV Resolution
0.1 mV || 0.1 mV
600.0 mV
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3) || Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3) || Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3) || Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)
6.000 V Resolution
0.001 V || 0.001 V
6.000 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3) || Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3) || Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3) || Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)
60.00 V Resolution
0.01 V || 0.01 V
60.00 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3) || Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3) || Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3) || Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)
600.0 V Resolution
0.1 V || 0.1 V
600.0 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3) || Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3)¹ || Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3) || Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)
1000 V Resolution
1 V || 1 V
1000 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(2% + 3) || Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3)¹ || Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(2% + 3) || Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)

Output Voltage

50 V (0% to +20%) Display range
0.01 to 6.00 MΩ || 6.0 to 50.0 MΩ
50 V (0% to +20%) Resolution
0.01 MΩ || 0.1 MΩ
50 V (0% to +20%) Test current
1 mA @ 50 kΩ || 1 mA @ 50 kΩ
50 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts)
±(3% + 5 counts) || ±(3% + 5 counts)
100 V (0% to +20%) Display range
0.01 to 6.00 MΩ || 6.0 to 60.0 MΩ || 60 to 100 MΩ
100 V (0% to +20%) Resolution
0.01 MΩ || 0.1 MΩ || 1 MΩ
100 V (0% to +20%) Test current
1 mA @ 100 kΩ || 1 mA @ 100 kΩ || 1 mA @ 100 kΩ
100 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts)
±(3% + 5 counts) || ±(3% + 5 counts) || ±(3% + 5 counts)
250 V (0% to +20%) Display range
0.1 to 60.0 MΩ  || 60 to 250 MΩ
250 V (0% to +20%) Resolution
0.1 MΩ || 1 MΩ
250 V (0% to +20%) Test current
1 mA @ 250 kΩ  || 1 mA @ 250 kΩ
250 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts)
±(1.5% + 5 counts) || ±(1.5% + 5 counts)
500 V (0% to +20%) Display range
0.1 to 60.0 MΩ || 60 to 500 MΩ
500 V (0% to +20%) Resolution
0.1 MΩ || 1 MΩ
500 V (0% to +20%) Test current
1 mA @ 500 kΩ  || 1 mA @ 500 kΩ
500 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts)
±(1.5% + 5 counts) || ±(1.5% + 5 counts)
1000 V (0% to +20%) Display range
0.1 to 60.0 MΩ  || 60 to 600 MΩ  || 0.6 to 2.0 GΩ
1000 V (0% to +20%) Resolution
0.1 MΩ || 1 MΩ || 100 MΩ
1000 V (0% to +20%) Test current
1 mA @ 1 MΩ || 1 mA @ 1 MΩ || 1 mA @ 1 MΩ
1000 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts)
±(1.5% + 5 counts) || ±(1.5% + 5 counts) || ±(10% + 3 counts)

Documents

Data Sheet

Download