Fluke

Fluke 1738/INTL

Fluke 1738/INTL
Fluke 1738/INTL
Fluke 1738/INTL
Fluke 1738/INTL
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
32 977,88 zł
1
Payment methods:

Current Inputs

Number of inputs
4, mode selected automatically for attached sensor
Input voltage
Clamp input: 500 mVrms/50 mVrms; CF 2.8
Rogowski coil input
150 mVrms/15 mVrms at 50 Hz, 180 mVrms/18 mVrms at 60 Hz; CF 4; all at nominal probe range
Range
40 mA to 4 A/0.4 A to 40 A with 40 A clamp i40s-EL
Bandwidth (-3 dB)
42.5 Hz - 3.5 kHz
Scaling
1:1 and variable

Wireless Connection

Number of inputs
2
Supported modules
Fluke Connect® 3000 series
Acquisition
1 reading/s

Data Acquisition

Resolution
16-bit synchronous sampling
Sampling frequency
10.24 kHz at 50/60 Hz, synchronized to mains frequency
Input signal frequency
50/60 Hz (42.5 to 69 Hz)
Circuit types
1-φ, 1-φ IT, Split phase, 3-φ delta, 3-φ wye, 3-φ wye IT, 3-φ wye balanced, 3-φ Aron/Blondel (2-element delta), 3-φ delta open leg, Currents only (load studies)
Data storage
Internal flash memory (not user replaceable)
Memory size
Typical 10 logging sessions of 8 weeks with 1-minute intervals and 500 events¹

Auxiliary Inputs

Number of inputs
2
Input range
0 to ±10 V dc, 1reading/s
Scale factor
Format: mx + b (gain and offset) user configurable
Displayed units
User configurable (7 characters, for example, °C, psi, or m/s)

Standards Compliance

Harmonics
IEEE 519 (short time harmonics)
Power quality
IEC 61000-4-30 Class S, IEC62586-1 (PQI-S device)
Power
IEEE 1459
Power quality compliance
EN50160 (for measured parameters)

Voltage Inputs

Number of inputs
4 (3 phases and neutral)
Maximum input voltage
1000 Vrms, CF 1.7
Input impedance
10 MΩ
Bandwidth (-3 dB)
42.5 Hz - 3.5 kHz
Scaling
1:1 and variable
Measurement category
1000 V CAT III/600 V CAT IV

Basic Interval

Measured parameters
Voltage, current, aux, frequency, THD V, THD A, power, power factor, fundamental power, DPF, energy
Averaging interval
User selectable: 1 sec, 5 sec, 10 sec, 30 sec, 1 min, 5 min, 10 min, 15 min, 30 min
Averaging time min/max values
Voltage, Current: Full cycle RMS updated every half cycle (URMS1/2 according to IEC61000-4-30 Aux, Power: 200ms

Interfaces

USB-A
File transfer via USB flash drive, firmware updatesMax. current: 120 mA
WiFi
File transfer and remote control via direct connection or WiFi infrastructure
Bluetooth
Read auxiliary measurement data from Fluke Connect® 3000 series modules (requires 1738 or 1736 upgrade option)
USB-mini
Data download device to PC

Environmental Specifications

Operating temperature
-10 °C to +50 °C (14 °F to 122 °F)
Storage temperature
-20 °C to +60 °C (-4 °F to 140 °F), with battery: -20 °C to +50 °C (-4 °F to 122 °F)
Operating humidity
40 °C to 50 °C (104 °F to 122 °F) max. 45% RH
Operating altitude
2000 m (up to 4000 m derate to 1000 V CAT II/600 V CAT III/300 V CAT IV)
Storage altitude
12,000m
Enclosure
IP50 in accordance with EN60529
Vibration
MIL-T-28800E, Type 3, Class III, Style B
Safety
IEC 61010-2-031: CAT IV 600 V / CAT III 1000 V
Electromagnetic compatibility
USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103
Temperature coefficient
0.1 x accuracy specification/°C

Power Supply

Voltage range
100 V to 240 V using standard power cord (IEC 60320 C7)
Power consumption
Maximum 50 VA (max. 15 VA when powered using IEC 60320 input)
Efficiency
≥ 68.2% (in accordance with energy efficiency regulations)
Maximum no-load consumption
< 0.3 W only when powered using IEC 60320 input
Mains power frequency
50/60 Hz ±15%
Battery
Li-ion 3.7 V, 9.25 Wh, customer-replaceable
On-battery runtime
Four hours in standard operating mode, up to 5.5 hours in power saving mode
Charging time
< 6 hours

Power Quality Measurements

Measured parameter
Voltage, frequency, unbalance, voltage harmonics, THD V, current, harmonics, THD A, TDD
Averaging interval
10 min
Individual harmonics
2nd to 50th harmonic
Total harmonic distortion
Calculated on 50 harmonics
Events
Voltage: dips, swells, interruptions, current: inrush current
Triggered recordings
Waveform of voltage and current (1738)

Demand Interval (Energy Meter Mode)

Measured parameters
Energy (Wh, varh, VAh), PF, maximum demand, cost of energy
Interval
User selectable: 5 min, 10 min, 15 min, 20 min, 30 min, off

General Specifications

Color LCD display
4.3-inch active matrix TFT, 480 pixels x 272 pixels, resistive touch panel
Warranty
One-year
Dimensions
Instrument with power supply attached: 19.8 cm x 16.7 cm x 9 cm (7.8 in x 6.6 in x 3.5 in)
Weight
115 g
Tamper protection
Kensington lock slot
Measuring range
1 to 150 A ac/10 to 1500 A ac
Nondestructive current
100 kA (50/60 Hz)
Intrinsic error at reference condition*
±0.7% of reading
Accuracy 173x + iFlex
±(1% of reading + 0.02% of range)
Temperature coefficient over operating temperature range
0.05% of reading/°C 0.09% of reading/°F
Working voltage
1000 V CAT III, 600 V CAT IV
Probe cable length
305 mm (12 in)
Probe cable diameter
7.5 mm (0.3 in)
Minimum bending radius
38 mm (1.5 in)
Output cable length
2 m (6.6 ft)
Probe cable material
TPR
Coupling material
POM + ABS/PC
Output cable
TPR/PVC
Operating temperature
-20 °C to +70 °C (-4 °F to 158 °F) temperature of conductor under test shall not exceed 80 °C (176 °F)
Temperature, non-operating
-40 °C to +80 °C (-40 °F to 176 °F)
Relative humidity, operating
15% to 85% non-condensing
IP rating
IEC 60529:IP50

Documents

Data Sheet

Download