Fluke 1750-TF

Fluke 1750-TF
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

Power Measurements

Power, battery life Measurement type
True RMS calculated continuously: every cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by standards
Frequency Measurement range
42.5 to 69 Hz
Frequency Measurement source
Same as PLL synchronization source
Frequency Measurement accuracy
±10 mHz (10 to 110% of range, with sine wave)
Power factor Measurement range
0.000 to 1.000
Power factor Measurement accuracy
±1 digit from the calculation of each measured value (±3 digits for total)
Displacement power factor Measurement method
Calculated from the phase difference between voltage fundamental and current fundamental
Displacement power factor Measurement range
1.000 (leading) to + 1.000 (lagging)
Displacement power factor Measurement accuracy
±0.5% reading ±2% full scale ±1 digit
Voltage unbalance and phase sequence Measurement method
Positive sequence voltage divided by negative sequence voltage, per IEC 61000-4-30
Harmonic voltage and current
Analysis window: rectangular
Harmonic voltage and current Analysis order
1st to 50th order
Harmonic voltage and current Measurement accuracy
Voltage / Current: 1st to 20th orders: ±0.5% reading ±0.2% full scale, 21st to 50th orders: ±1% reading ±0.3% full scale (current sensor accuracy must be included for current and power)
Harmonic voltage and current Measurement method
IEC 61000-4-7
Inter-harmonic voltage and current (intermediate harmonics)
Analysis window: rectangular
Inter-harmonic voltage and current (intermediate harmonics) Analysis orders
1.5 to 49.5th order
Inter-harmonic voltage and current (intermediate harmonics) Measurement method
IEC 61000-4-7
Flicker Measurement method
IEC 61000-4-15 || Plt for 2 hours and Pst for 10 minutes
Flicker
Measuring range: 0, 1 to 5 (25) depending on voltage level, modulation and frequency

Technical Specifications

Power quality measurement standards Conformance
IEC 61999-1-4 Class 1, IEC 61000-4-30 Class A, S or B depending on measurement function, IEEE519, IEEE1159, IEEE1459 and EN50160
Power quality measurement standards Clock/calendar
Leap years, 24-hour clock
Power quality measurement standards Real-time clock accuracy
Not more than ±1 s/day
Power quality measurement standards Internal memory capacity for data
At least 2 GB
Power quality measurement standards
Maximum recording period: At least 31 days || Operating time during interruptions (internal UPS operation): 5 minutes per interruption, 60 minutes total operating time without recharging
Power quality measurement standards Measurement time control
Automatic
Power quality measurement standards Maximum number of events
Limited only by the size of the internal memory
Power quality measurement standards Power requirements
100 to 240 Vrms ±10%, 47-63 Hz, 40 W
Power quality measurement standards Dimensions
215 x 310 x 35 mm (8.5 x 12.2 x 3.5 in)
Power quality measurement standards Mass (weight)
6.3 kg (14 lb)
Input Measurement types
One Phase Plus Neutral, One Phase IT No Neutral, One Phase Split Phase, Three Phase Wye, Three Phase Delta, Three Phase IT, Three Phase High Leg, Three Phase Open Leg, 2 Element Delta, 21/2 Element Wye
Input Input channels
Voltage: 4 channels, AC/DC || Current: 5 channels
Input
Voltage channels: Input resistance: 2 MΩ || Voltage channels: Input capacitance: < 20 pF || Current input characteristics: 2 Vrms = full scale, 1 MΩ Input Impedance for ferro CTs, low impedance for Flexi-CTs || Measuring method: Simultaneous digital sampling of voltage and current. Digital PLL synchronized sampling, internal frequency reference used during voltage drops.
Synchronization and sampling PLL-synchronization source
The PLL synchronizes to the A-N voltage for wye power types, and to the A-B voltage for delta power types. All listed power types can be characterized as either wye or delta.
Synchronization and sampling PLL lock range
42.5 to 69 Hz
Synchronization and sampling
Sampling frequency: Voltage and current: 256 samples/cycle Inter-harmonics per IEC 61000-4-7: 2560 points / 10 cycles (50 Hz), 3072 points / 12 cycles (60 Hz) Transient Voltage: 5 MHz
Synchronization and sampling A/D resolution
Voltage and current: 24 bits || Transient voltage: 14 bits
Voltage and current measurements
Voltage measurement range: AC voltage: 1000 Vrms ±10% over range || Voltage measurement range: DC voltage: ±1000 V + 10% over range || Voltage crest factor: 3 or less
Voltage and current measurements Current measurement range
Depends on current probe used
Voltage and current measurements Current crest factor
4 or less
RMS voltage Measurement type
True RMS calculated continuously: every cycle, every 1/2 cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by IEC 61000-4-30.
RMS voltage
Measurement uncertainty: AC: ±0.2% reading ±0.1% full scale, above 50 Vrms || Measurement uncertainty: DC: ±0.5% reading ±0.2% full scale, above 50 VDC
RMS current Measurement type
True RMS calculated continuously: every cycle, every 1/2 cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by standards
Transient voltage (impulse) Measurement type
Waveshape sampling
Transient voltage (impulse) Full scale
8000 V pk
Transient voltage (impulse) Sample resolution
200 nS
Transient voltage (impulse)
Measurement uncertainty: ±5% reading ±20 V (test parameters: 1000 VDC, 1000 Vrms, 100 kHz)
Voltage swell (rms swell) Measurement type
True RMS (one cycle calculation by overlapping each half cycle - voltage between lines is measured for 3P3W lines and phase voltage is measured for 3P4W lines)
Voltage swell (rms swell) Displayed data
Amplitude and duration of swell
Voltage swell (rms swell) Measurement
Same as rms voltage
Voltage dip (rms sag) Measurement type
True RMS (one cycle calculation by overlapping each half cycle - voltage between lines is measured for 3P3W lines and phase voltage is measured for 3P4W lines)
Voltage dip (rms sag) Displayed data
Amplitude and duration of dip or interruption
Voltage dip (rms sag) Measurement
Same as rms voltage
Voltage dropout (interruption) Measurement type
Same as voltage dip
LAN interface Connector
RJ-45
LAN interface Speed and type
10/100 Base-T, auto MDIX
LAN interface
Communications protocol: TCP/IP over Ethernet
Wireless controller interface Connection
wireless (2.4 GHz radio)
Wireless controller interface Speed
up to 700 kbit/second
Wireless controller interface
Communications protocol: Bluetooth SPP

Environmental Specifications

Environmental
Operating environment: Indoors or in covered area outdoors, up to 2000 m altitude || Operating temperature and humidity: 0°C to 40°C, 80% RH max, non-condensing
Environmental Storage temperature and humidity
-20°C to 50°C, 80% RH max, non-condensing
Maximum rated working voltage
Voltage terminals: 1100 Vrms || Voltage durability: 5550 Vrms AC for 1 minute, between voltage input terminals, voltage input terminals and current probes, and voltage input terminals and case (50/60 Hz, 1 mA sense current)
Maximum rated working voltage Enclosure protection
IP30 (per EN 60529)
Standards EMC
EN 61326-1:1997+A1:1998 Class A || EN 61000-3-2:1995+A1:1998+A2:1998 || EN 61000-3-3:1995