Uncertainties

Uncertainty at reference conditions
Uncertainty including the voltage sensors is in compliance with IEC 61000-4-30 Class-A. All voltage sensors are suitable for DC up to 5 kHz || With Sensor 1000 V: 0,1% at Udin = 480 V and 600 V P-N || Sensor 600 V: 0,1% at Udin = 230 V P-N
Intrinsic uncertainty for harmonics
Class I as per EN 61000-4-7:2002
Reference conditions
23°C ±2 K < 60% rH; 74°F ±2 K < 60% rH || Warmed up instrument > 3h || Power supply: 100 V to 250 VAC
Temperature drift
100 ppm / K || Change of amplification through temperature: < 0.005% / K
Aging
< 0.05% / year || Change of amplification due to ageing: < 0.04% / year
Common mode rejection
> 100 dB at 50 Hz
Noise
Noise voltage, input short-circuited: < 40 µV
DC
±(0.2% rdg + 0.1% sensor)

General Data

Intrinsic uncertainty
refers to reference conditions and is guaranteed for two years
Quality system
developed, manufactured as per ISO 9001: 2000
Environment conditions
Operating temp. range: 0°C to +50°C; 32°F to +122°F || Working temp. range: -20°C to +50°C; -4°F to +122°F || Max. operating altitude: 2000 m: max. 600 V CAT IV1, power supply: 300 V CAT III 5000 m: max 600 V CAT III1, power supply: 300 V CAT II || Max. operating altitude: 1. Depending on sensor
Environment conditions Storage temp. range
20°C to +60°C; -4°F to 140°F
Environment conditions Reference temperature
23°C ± 2 K; 74°F ± 2 K
Environment conditions Climatic class
B2 (IEC 654-1), -20°C to +50°C; -4°F to +122°F
Reference conditions Environment temp.
23°C ± 2 K; 74°F ± 2 K
Reference conditions Power supply
230 V ± 10%
Reference conditions Power frequency
50 Hz / 60 Hz
Reference conditions Signal
declared input voltage Udin
Reference conditions
Averaging: 10 minute intervals
Housing
Insulated, robust plastics housing
EMC Emission
Class-A as per IEC/EN 61326-1
EMC
Immunity: IEC/EN 61326-1
Power supply Range
AC: 83 V to 264 V, 45 to 65 Hz || DC: 100 V to 375 V
Power supply Safety
IEC/EN 61010-1 2nd edition || 300 V CAT III
Power supply Power consumption
NIMH, 7.2 V, 2.7 Ah
Power supply Battery pack
In case of a power supply failure an internal battery maintains the supply for up to 40 minutes. Afterwards, or in case of discharged accumulators the Fluke 1760 is turned off and continues the measurements with the latest settings as soon as the supply voltage returns. The battery can be replaced by the user.
Display
Fluke 1760 features LED indicators for the status of the 8 channels, phase sequence, power supply (mains or accumulator), memory usage, time synchronization, and data transfer.
Display Power LED
Permanent light: normal power supply from mains.OFF: supply via internal accumulator in case of a power failure.
Display Channel LEDs
3-color LEDs per channel for:Overload conditionUnder load condition
Data memory
2 GB Flash memory depending on model
Memory model
Linear
Interfaces
Ethernet (100 MB/s), compatible to Windows® 98/ME/NT/2000/XP RS 232, external modem via RS 232
Baud rate for RS 232
9600 Baud to 115 kBaud
Dimensions (H x W x D)
325 x 300 x 65 mm (2.8 x 11.8 x 2.6 in)
Weight
appr. 4.9 kg; 10.8 lbs (without accessories)
Warranty
2 years
Calibration interval
1 year recommended for Class-A, otherwise 2 years

Signal Conditioning

Range for 50 Hz systems
50 Hz ±15% (42.5 Hz to 57.5 Hz)
Range for 60 Hz systems
60 Hz ±15% (51 Hz to 69 Hz)
Resolution
16 ppm
Sampling frequency for 50 Hz power frequency
10.24 kHz, The sampling rate is synchronized to mains frequency.
Uncertainty for frequency measurements
< 20 ppm
Uncertainty of internal clock
< 1 s/day
Measurement intervals
Aggregation of the interval values as per IEC 61000-4-30 Class-A || Min-, Max-values: Half cycle, e.g.: 10 ms rms values at 50 Hz
Measurement intervals Transients
Sample rate 100 kHz to 10 MHz per channel
Harmonics
as per IEC 61000-4-7:2002: 200 ms
Flicker
as per EN 61000-4-15:2003: 10 min (Pst), 2 h (Plt)

Overview of Measurement Functions

Statistical evaluation
Power quality statistics according to EN50160 and DISDIP tables like ITIC, CEBEMA, ANSI
Event list
Dips, swells and interruptions are detected and stored in the event list. Also any trigger which fires generates an event added to this list. || The Event list shows the exact time when the event occurred as well as the duration and magnitude. Sorting by several attributes of these events is possible to select one for further root cause analysis. || RMS values, transients and fast transients can be stored if a trigger fires
Continuous recording
Fluke 1760 records rms values together with corresponding minimum and maximum values for:• Voltage• Current• Power P, Q, S• Power Factor• kWh• Flicker• Unbalance• Frequency• Harmonics/InterharmonicsContinuously with the following time aggregations:Day10 minFree Interval, e.g.: 15 min, 2 h
Triggered recordings rms
Aggregation time is adjustable between 10 ms (1/2 cycle), 20 ms (1 cycle), 200 ms (10/12 cycles) or 3 sec (150/180 cycles). || Calculating rms values, Harmonics and Interharmonics is performed synchronous to the power frequency. || Basic aggregation for harmonics and interharmonics is 200ms
Triggered recordings Oscilloscope:
Sample rate is 10.24 kHz for all 8 channels
Triggered recordings
Mains signalling: Phases and N-conductor, Voltage and current
Online mode
Variable refresh rate. This feature allows verification of instrument set up and delivers a quick overview of oscilloscope, transients and events.

Measurement Inputs

Number of inputs
8 galvanically isolated inputs for voltage and current measurements.
Sensor safety
up to 600 V CAT IV depending on sensor
Basic safety
300 V CAT III
Nominal voltage (rms)
100 mV
Range (peak value)
280 mV
Overload capacity (rms)
1000 V, continuously
Voltage rise rate
max. 15 kV/µs
Input resistance
1 MΩ
Input capacitance
5 pF
Input filter
Each channel is equipped with a passive low-pass filter, an anti-aliasing filter and a 16-bit A/D converter. All channels are sampled synchronously with a common quartz-controlled clock pulse.The filters protect against voltage transients and limit the signal rise rate, reduce high frequency components and especially the noise voltage above half the sampling rate of the A/D converter by 80 dB, thus achieving very small measuring errors in an exceptionally large amplitude range. This is also valid under extreme operating conditions like transient voltages at the output of converters.