Safety and Environmental Specifications

General safety
IEC 61010-1: Pollution Degree 2
Measurement safety
IEC 61010-2-032: CAT III 600 V / CAT IV 300 V
Operating temperature
-10 °C to +50 °C
Storage temperature
-40 °C to +60 °C
Operating humidity
Non-condensing (<10 °C) || 90% relative humidity (10 °C to 30 °C) || 75% relative humidity (30 °C to 40 °C) || 45% relative humidity (40 °C to 50 °C)
Protection class
IEC 60529: IP30 (jaw closed)
Operating altitude
2000 m
Storage altitude
12000 m
Current sensor action category
IEC 61557-13: Class 1, ≤100 A / m

Physical

Display (LCD)
Digital readout: 3300 count
Display refresh rate
4 times / sec
Maximum conductor diameter
61 mm
Dimensions
257 x 116 x 46 mm
Weight
600 g
Battery
2 AA, IEC LR6, NEDA 15A, alkaline
Battery life
Expected battery life, without using backlight and spotlight, is more than 150 hours
Automatic shut-down
Meter automatically shuts down after 15 minutes of inactivity

Electrical Specifications

Measurement function
Alternating current
AC ampere range
3 mA, 30 mA, 300 mA, 3 A, 30 A, 60 A
Range selection
mA / A: manual selection3 mA / 30 mA / 300 mA: automatic selection3 A / 30 A / 60 A: automatic selection
Range / Resolution
3 mA¹ / 0.001 mA || 30 mA / 0.01 mA || 300 mA / 0.1 mA || 3 A / 0.001 A || 30 A / 0.01 A || 60 A / 0.1 A
Accuracy – filter activated (40 to 70 Hz), filter off (40 to 1 kHz)²
3 mA – 30 A60 A: 1% + 5 digits2% + 5 digits
Frequency
40 Hz to 1 kHz
Crest Factor
3

International

IEC 61326-1
Industrial Electromagnetic Environment
CISPR 11
Group 1 Class B
Group 1
Generated inside the equipment and / or use radio frequency energy associated with conducting the energy for the device's own internal functions is essential.
Class B
Equipment for home appliances and residential buildings directly connected to a low voltage power supply network equipment. When this device is connected to a test object, it may exceed the emission levels produced by CISPR 11 requirements.Korea (KCC): Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A
This product meets the requirements of industrial electromagnetic equipment, vendors or users should be aware of this. This device is intended for use in a commercial environment, instead of the home environment.USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.