- Built-in reference thermometer readout accuracy (4-wire reference probe)3
- ±0.010°C at –95°C
- Built-in reference thermometer readout accuracy (4-wire reference probe)
- ±0.025°C at 140°C
- Reference resistance range
- 0 Ω to 400 Ω
- Reference resistance accuracy4
- 0 Ω to 42 Ω: ±0.0025 Ω42 Ω to 400 Ω: ±60 ppm of reading
- Reference characterizations
- ITS-90, CVD, IEC-751, resistance
- Reference measurement capability
- 4 wire
- Reference probe connection
- 6-pin din with INFO-CON technology
- Built-in RTD thermometer readout accuracy
- PT-100 (JIS): ±0.02°C at 0°C
- RTD resistance range
- 0 Ω to 400 Ω
- Resistance accuracy4
- 0 Ω to 25 Ω: ±0.002 Ω
- Resistance accuracy
- 25 Ω to 400 Ω: ±80 ppm of reading
- RTD characterizations
- PT-100 (385), (JIS), (3926), NI-120, resistance
- RTD measurement capability
- 2-wire, 3-wire, and 4-wire RTD with jumpers only
- RTD connection
- 4-terminal input
- Built-in TC thermometer readout accuracy5
- Type J: ±0.70°C at 140°C
- Built-in TC thermometer readout accuracy
- Type C: ±1.00°C at 140°C
- TC millivolt range
- –10 mV to 75 mV
- Voltage accuracy
- 0.025% of reading +0.01 mV
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- TC connection
- Miniature connectors (ASTM E1684)
- Built-in mA readout accuracy
- 0.02% of reading + 0.002 mA
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- MA connection
- 2 terminal input
- Loop power function
- 24 V DC loop power
- Built-in electronics temperature coefficient(0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per °C