Thermistor Equivalent Temperature Accuracy (Readout Only)

100°C
±0.030
50°C
±0.006
25°C
±0.003
75°C
±0.014
0°C
±0.002

Thermocouple Equivalent Temperature Accuracies (Readout Only)

Type K
±0.61°C from -200°C to 0°C ±0.24°C from 0°C to 1370°C
Type U
±0.54°C from -200°C to 0°C ±0.24°C from 0°C to 400°C
Type B
±0.85°C from 600°C to 800°C ±0.68°C from 800°C to 1000°C±0.57°C from 1000°C to 1800°C
Type T
±0.60°C from -200°C to 0°C ±0.25°C from 0°C to 400°C
Type C
±0.32°C from 100ºC to 550°C ±0.71°C from 550°C to 2300°C
Type R
±1.09°C from -20°C to 0°C ±0.97°C from 0°C to 500°C±0.49°C from 500°C to 1750°C
Type U
Accuracies are based on internal Reference Junction Compensation. Refer to Technical manual for equivalent accuracies with an external reference junction.
Type L
±0.36°C from -200°C to 0°C ±0.23°C from 0°C to 1370°C
Type J
±0.52°C from -200°C to 0°C ±0.23°C from 0°C to 1200°C
Type S
±1.05°C from -20°C to 0°C ±0.95°C from 0°C to 500°C±0.56°C from 500°C to 1750°C
Type N
±0.72°C from -200°C to 0°C ±0.28°C from 0°C to 1300°C
Type E
±0.52°C from -200°C to 0°C±0.22°C from 0°C to 950°C
Type M
±0.26°C from -20°C to 0°C ±0.25°C from 0°C to 400°C±0.22°C from 400°C to 1400°C

PRT Equivalent Temperature Accuracy (Readout Only)

–100°C
±0.011
100°C
±0.019
0°C
±0.015
600°C
±0.039
200°C
±0.023
400°C
±0.031

Specifications

Temperature coefficient, voltage (–10°C to 13°C , +33°C to 60°C)
±(0.001%/°C + 1 μV/°C)
Millivolt range and accuracy
–10 mV to 75 mV ±(0.005% + 5 μV)
Temperature coefficient, resistance (–10°C to 13°C , +33°C to 60°C)
0.0008%/°C + 0.0004 Ω (0 Ω to 400 Ω) 0.002%/°C + 0.1 Ω (0 Ω to 50 kΩ) 0.06%/°C + 0.1 Ω (50 kΩ to 500 kΩ)
Excitation current, resistance
1 mA (0 Ω to 400 Ω) 10 μA (0 Ω to 50 kΩ) 2 µA (50 kΩ to 500 kΩ)
Weight
0.65 kg (1.4 lb)
Sample interval(normal)
1 second
Logging
25 readings with statistics
Sensor types
PRTs, RTDs, Thermistors, and Thermocouples
Size
96 x 200 x 47 mm (3.75 x 7.9 x 1.86 in)
Sample interval (fast mode)
0.3 seconds (see technical manual for details)
Input channels
1
Environmental conditions for best accuracy
13°C to 33°C
Power requirements
3 AA alkaline batteries
Operating temperature
-10°C to 60°C (Best accuracy 13°C to 33°C)
Resistance range and accuracy
0 Ω to 400 Ω ±(0.004% + 0.002 Ω) 200 Ω to 50 kΩ ±(0.01% + 0.5 Ω) 50 kΩ to 500 kΩ ±(0.03%)
Thermocouple types
C,E,J,K,L,M,N,T,U,B,R,S

Accuracies with Selected Probes (±°C)

100°C
5616-12: 0.027 5615-6: 0.028 5627A-12: 0.065 5610-9: 0.009
0°C
5616-12: 0.021 5615-6: 0.021 5627A-12: 0.049 5610-9: 0.009
-200°C
5616-12: 0.014 5615-6: 0.025 5627A-12: 0.027 5610-9: n/a
420°C
5616-12: 0.050 5615-6: n/a 5627A-12: 0.130 5610-9: n/a
420°C
Includes readout accuracy, probe calibration, and probe drift
300°C
5616-12: 0.040 5615-6: 0.043 5627A-12: 0.103 5610-9: n/a