Fluke

9190A-E

1 / 4
thumbnail
thumbnail
thumbnail
thumbnail
Part Number
9190A-E
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

P Specifications

Reference probe connection
6-pin din with INFO-CON technology
RTD resistance range
0 Ω to 400 Ω
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
1. For ambient temperature of 23°C. 2. Time from when the SETPOINT is reached to when the unit is with in stability specification. 3. The temperature range may be limited by the reference probe connected to the readout. The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 4. Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05 Ω to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires. 5. The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz.
Built-in TC thermometer readout accuracy5
Type J: ±0.70°C at 140°C
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.013°C at –25°C
Built-in RTD thermometer readout accuracy
PT-100 (JIS): ±0.02°C at 0°C
Voltage accuracy
0.025% of reading +0.01 mV
TC connection
Miniature connectors (ASTM E1684)
MA connection
2 terminal input
Built-in TC thermometer readout accuracy5
Type T: ±0.60°C at 140°C
Built-in TC thermometer readout accuracy5
Type R: ±1.60°C at 140°C
Reference resistance range
0 Ω to 400 Ω
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.020°C at 50°C
TC millivolt range
–10 mV to 75 mV
RTD characterizations
PT-100 (385), (JIS), (3926), NI-120, resistance
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C
Built-in TC thermometer readout accuracy5
Type M: ±0.65°C at 140°C
Built-in mA readout accuracy
0.02% of reading + 0.002 mA
Reference measurement capability
4 wire
Built-in TC thermometer readout accuracy5
Type L: ±0.65°C at 140°C
Built-in TC thermometer readout accuracy5
Type N: ±0.75°C at 140°C
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.015°C at 0°C
Built-in TC thermometer readout accuracy5
Type E: ±0.60°C at 140°C
MA range
Cal 4-22 mA, Spec 4-24 mA
Loop power function
24 V DC loop power
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C
Built-in TC thermometer readout accuracy5
Type K: ±0.75°C at 140°C
Resistance accuracy4
25 Ω to 400 Ω: ±80 ppm of reading
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Reference resistance accuracy4
0 Ω to 42 Ω: ±0.0025 Ω 42 Ω to 400 Ω: ±60 ppm of reading
Built-in TC thermometer readout accuracy5
Type C: ±1.00°C at 140°C
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.025°C at 140°C
Built-in RTD thermometer readout accuracy
PT-100 (385): ±0.02°C at 0°C
RTD connection
4-terminal input
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.010°C at –95°C
Reference characterizations
ITS-90, CVD, IEC-751, resistance
Resistance accuracy4
0 Ω to 25 Ω: ±0.002 Ω
Built-in RTD thermometer readout accuracy
PT-100 (3926): ±0.02°C at 0°C
RTD measurement capability
2-wire, 3-wire, and 4-wire RTD with jumpers only
Built-in TC thermometer readout accuracy5
Type U: ±0.70°C at 140°C
Built-in TC thermometer readout accuracy5
Type S: ±1.60°C at 140°C

Specifications

Display accuracy
±0.2°C full range
Loading effect
(versus display with one 6.35 mm probe)
Electromagnetic environment
IEC 61326-1: Basic
Loading effect
(with a 6.35 mm reference probe and three 6.35 mm probes)
Stabilization time2
15 min
Operating conditions
0°C to 35°C, 0% to 90%
Cooling time1
140°C to 23°C: 60 min
Computer interface
RS-232, USB Serial, and 9930 interface-it temperature calibration software included
Immersion (well) depth
160 mm (6.3 in)
Radial gradient
±0.01°C full range
Loading effect
±0.25°C at –95°C
Refrigerants
R32 (Difluoromethane)
Power requirements
100 V to 115 V (±10%) 50/60 Hz, 575 W
Stability
±0.015°C full range
Accuracy with external reference3
±0.05°C full range
Environmental conditions for all specifications except temperature range
13°C to 33°C
Axial uniformity at 40 mm (1.6 in)
±0.05°C full range
Cooling time1
23°C to –90°C: 80 min
Cooling time1
23°C to –95°C: 90 min
Display
LCD, °C or °F user selectable
Safety
IEC 61010-1, Installation Category II, Pollution degree 2
Heating time1
–95°C to 140°C: 40 min
Refrigerants
< 20 g, ASHRAE safety group A2L
Power requirements
200 V to 230 V (±10%) 50/60 Hz, 575 W
Weight
16 kg (35 lb)
Loading effect
±0.006°C full range
Resolution
0.01°
Refrigerants
R704 (Helium)
Operating conditions
RH (non-condensing) < 2000 m altitude
Well diameter
30 mm (1.18 in)
Temperature range at 23°C
–95°C to 140°C (–139°F to 284°F)
System fuse ratings
115 V: 6.3 A T 250 V
Refrigerants
< 20 g, ASHRAE safety group A1
Loading effect
±0.10°C at 140°C
Size (H x W x D)
480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in)
4–20 mA fuse (-P model only)
50 mA F 250 V
System fuse ratings
230 V: 3.15 A T 250 V