General Specifications

Measuring functions
3-pole earth ground resistance, 2-pole ac resistance of a conductor, Interference voltage
Intrinsic error
Refers to the reference temperature range and is guaranteed for one year
Measuring rate
2 measurements/second
Battery1
One 9 volt alkaline (LR61)
Battery condition
LO-BAT is displayed if voltage drops below 6.5 V
Voltages Between jacks H/C2 and E/C1
250 Veff maximum (effective voltage)
Voltages Between jacks S/P2 and E/C1
250 Veff maximum
Climatic class
VDE/VDI 3540 RZ (conforming to KWG as per DIN 40040, 4/87)
Temperature performance2
Working: -10°C to +50°C (+14°F to +122°F)
Temperature performance
Operating: 0°C to +35°C (+32°F to +95°F)
Temperature performance Storage
-20°C to +60°C (+68°F to +140°F)
Temperature performance Reference
+23°C ±2°C (+73°F ±4°F)
Temperature coefficient
±0.1% of range per degree Kelvin
Safety
IEC/EN 61010-1, 600 V CAT II, pollution degree 2
Dimensions
113 x 54 x 216 mm (4.5 x 2.1 x 8.5 in), including holster
Weight
850 g (1.9 lb), including standard accessories, volume approximately 600 cm3

Measuring Range

0.15 Ω to 20 Ω Resolution
0.01 Ω
0.15 Ω to 20 Ω Display range
0 to 19.99 Ω
200 Ω Resolution
0.1 Ω
200 Ω Display range
20 to 199.9 Ω
2 kΩ Resolution
1 Ω
2 kΩ Display range
200 to 1999 Ω
Intrinsic Uncertainty
±(6% of measured value + 5D)
Operating uncertainty IEC 615573
±(18% of measured value + 5D)

Electrical Specifications

Maximum deviations E1 Influence factor
Position
Maximum deviations
E1 Deviation influence: 0% || E2 Deviation influence: 0% || E3 Deviation influence: 2.3% || E4 Deviation influence: 0.6% || E5 Deviation influence: 10%
Maximum deviations E2 Influence factor
Supply voltage
Maximum deviations E3 Influence factor
Temperature E3
Maximum deviations E4 Influence factor
Serial interference voltage (20 V)
Maximum deviations E5 Influence factor
Probe- and auxiliary probe resistance
Test voltage
3.7 kV
Protection type
IP 40; IEC/EN 60529
Electromagnetic compatibility Emission
IEC/EN 61326 Class B
Electromagnetic compatibility
Immunity: IEC/EN 61326 Annex C
RE Resistance measurement
Measuring method: Current-voltage measurement with improved cross-talk attenuation, no compensation of measuring lead resistance, with probe (3-pole) or without probe (2-pole), as per IEC/EN 61557–5
R Open circuit voltage
23 to 24 V AC
R Short circuit current
> 50 mA AC
R
Measuring frequency: 128 Hz
R Maximum permissible overload
250 Veff
Measuring time
8 seconds (average from when START is pressed)
Limit input
Tester retains set value even if instrument is turned off (assuming battery power supply is sufficient)
Automatic changeover of resolution RH
< 7 kΩ || < 50 kΩ || > 50 kΩ
Automatic changeover of resolution Resolution
0.01 Ω || 0.1 Ω || 1 Ω
Interference voltage display DC + AC
Vmax: 30 Veff || Common mode rejection: > 80 dB at 50 Hz and 60 Hz || Measuring uncertainty: < 10% for pure AC and DC signals
Interference voltage display DC + AC Ri
680 kΩ