Description
Active Mature
- Country of Origin
- CH
- Part Number
- XMLPM00GC21F
- Tariff Code
- 90262020
- EAN
- 3389119624916
- UNSPSC
- 39122228
Payment methods:
- Best offer from:
- Trusted Automa.Net Supplier
- Availability:
- 1 pc.In-StockUpdated at: 21-10-2024
- Estimated Delivery:
- On Request
- Condition:
- New Surplus
New Surplus - A new, unused, undamaged item without a manufacturer guarantee. The item may come without original packaging or in replacement packaging. If provided, the box may have been opened.
Secure Payment Process
Return & Refund Policies
Shipping Policies
Warranty Policies
Specifications
- Pressure Sensor Type
- pressure transmitter
- Type Of Output Signal
- analogue
- Range Of Product
- Telemecanique Pressure sensors XM
- Electrical Connection
- 1 male connector EN 175301-803-A (ex DIN43650) 3 pins
- Pressure Sensor Size
- -1-0 bar
- Product Or Component Type
- electronic pressure sensors
- Type Of Packing
- individual
- Local Display
- without
- Pressure Sensor Name
- XMLP
- Current Consumption
- < 23 mA
- Fluid Connection Type
- G 1/4A (male) DIN 3852-E
- Analogue Output Function
- 4...20 mA, 2-wire
- Controlled Fluid
- fresh water 0-125 °C) || air -15-125 °C) || gas -15-125 °C) || hydraulic oil -15-125 °C)
- Quantity Per Set
- set of 1
- Electrical Circuit Type
- control circuit
- [Us] Rated Supply Voltage
- 12...24 V DC SELV 7-33 V)
Complementary
- Accuracy
- 0.1 % of the measuring range
- Diameter
- 26 mm
- [Uimp] Rated Impulse Withstand Voltage
- 0.5 kV
- Repeat Accuracy
- +/- 0.25 % of the measuring range
- Drift Of The Zero Point
- +/- 0.02 % of measuring range/°K
- Electromagnetic Compatibility
- susceptibility to electromagnetic fields EN/IEC 61000-4-3 10 V/m 80...1000 MHz || radiated RF fields EN/IEC 61000-4-6 10 V 0.15...80 MHz || electrostatic discharge immunity test EN/IEC 61000-4-2 8 kV air, 4 kV contact || 1.2/50 µs shock waves immunity test EN/IEC 61000-4-5 1 kV 42 Ohm || immunity to magnetic fields EN/IEC 61000-4-8 100 A/m 50 Hz || electrical fast transient/burst immunity test EN/IEC 61000-4-4 4 kV
- Materials In Contact With Fluid
- stainless steel AISI 316L || ceramic || fluorocarbon FPM
- Length
- 55 mm
- Net Weight
- 0.095 kg
- Mechanical Durability
- 10000000 cycles
- Operating Position
- any position
- Measurement Accuracy
- +/- 0.5 % of the measuring range
- Protection Type
- load short-circuit || reverse polarity
- Drift Of The Sensitivity
- +/- 0.02 % of measuring range/°K
- Maximum Permissible Accidental Pressure
- 3 bar
- Destruction Pressure
- 5 bar
- Response Time On Output
- <= 2 ms 10...90 % of full scale
- Pressure Setting Range
- -1-0 bar
Offer Sustainability
- Rohs Exemption Information
- Yes
- Mercury Free
- Yes
- Environmental Disclosure
- ENVPEP1511013EN
- Sustainable Offer Status
- Green Premium product
- Reach Regulation
- Free of Substances of Very High Concern above the threshold
- California Proposition 65
- WARNING: This product can expose you to chemicals including: Diisononyl phthalate (DINP), which is known to the State of California to cause cancer, and Di-isodecyl phthalate (DIDP), which is known to the State of California to cause birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
- Reach Free Of Svhc
- Yes
- Eu Rohs Directive
- Pro-active compliance (Product out of EU RoHS legal scope)
Environment
- Product Certifications
- EAC || RCM || CE || cULus
- Ambient Air Temperature For Operation
- -30-85 °C
- Standards
- NSF ANSI 61 || EN/IEC 61326-2-3
- Shock Resistance
- 100 gn 11 ms EN/IEC 60068-2-27
- Vibration Resistance
- 20 gn 10-2000 Hz) EN/IEC 60068-2-6
- Ip Degree Of Protection
- IP65 conforming to EN/IEC 60529
- Ambient Air Temperature For Storage
- -50-100 °C