Fluke
9143-E-156
1 / 2
- Part Number
- 9143-E-156
- UPC
- 095969420068
- Get up to 90 days payment terms from 500+ suppliers
- Reselling discounts up to 40%!
Send request for price
1
Payment methods:
P specifications
- Built-in RTD thermometer readout accuracy
- NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- Built-in reference thermometer readout accuracy (4-Wire reference probe)1
- ±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
- RTD connection
- 4 terminal input
- MA connection
- 2 terminal input
- TC connection
- Small connectors
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- 1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
- Reference measurement capability
- 4-wire
- Voltage accuracy
- 0.025% of reading + 0.01 mV
- RTD measurement capability
- 4-wire RTD (2-,3-wire RTD w\ Jumpers only)
- RTD resistance range
- 0 ohms to 400 ohms
- Reference characterizations
- ITS-90, CVD, IEC-751, Resistance
- Reference resistance range
- 0 ohms to 400 ohms
- RTD resistance accuracy‡
- 0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- Reference resistance accuracy2
- 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
- Built-in mA readout accuracy
- 0.02% of reading + 2 mV
- Built-in TC thermometer readout accuracy
- Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
- RTD characterizations
- PT-100 (385),(JIS),(3926), NI-120, Resistance
- Reference probe connection
- 6-pin Din with Infocon Technology
- TC millivolt range
- –10 mV to 75 mV
- Loop power function
- 24 V DC loop power
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per °C
Base Unit Specifications
- Hysteresis
- 0.03
- Power requirements
- 100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
- Operating conditions
- 0°C to 50°C, 0% to 90% RH (non-condensing)
- Insert OD
- 25.3 mm (1.00 in)
- Cooling time
- 32 min: 350°C to 33°C 14 min: 350°C to 100°C
- Axial uniformity at 40 mm (1.6 in)
- ±0.04°C at 33°C ±0.1°C at 200°C±0.2°C at 350°C
- Weight
- 7.3 kg (16 lb)
- Display accuracy
- ±0.2°C Full Range
- Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
- ±0.015°C Full Range
- Computer interface
- RS-232 and 9930 Interface-it control software included
- Stability
- ±0.02°C at 33°C ±0.02°C at 200°C±0.03°C at 350°C
- Heating time
- 5 min: 33°C to 350°C
- Display
- LCD, °C or °F user-selectable
- Radial uniformity
- ±0.01°C at 33°C ±0.015°C at 200°C±0.02°C at 350°C
- Temperature range at 23°C
- 33°C to 350°C (91°F to 662°F)
- Size (H x W x D)
- 290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
- Immersion (well) depth
- 150 mm (5.9 in)
- Resolution
- 0.01°
- Environmental conditions (for all specifications except temperature range)
- 13°C to 33°C
Other Categories
ActuatorsBlocksCables, Pipes & HosesCompressorsControlCouplings & FittingsCylinders & PistonsElectronicsFiltersHVACIndustrial PC's & HMIsLightingLinear techniqueMotorsPower SupplyPower TransmissionPumps & ValvesRenewable energyRoboticsSeals, Rings & ConnectorsSensors & Measuring DevicesTools and AccessoriesCNCOther