P Specifications

Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Loop power function
24 V DC loop power
TC millivolt range
–10 mV to 75 mV
MA range
Cal 4-22 mA, Spec 4-24 mA
MA connection
2 terminal input
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
TC connection
Small connectors
Voltage accuracy
0.025% of reading + 0.01 mV
Reference resistance range
0 ohms to 400 ohms
Built-in reference thermometer readout accuracy (4-wire reference probe)1
±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
Reference measurement capability
4-wire
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
RTD resistance range
0 ohms to 400 ohms
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
Built-in mA readout accuracy
0.02% of reading + 2 mV
RTD resistance accuracy2
25 ohms to 400 ohms: ±80 ppm of reading 0 ohms to 25 ohms: ±0.002 ohms
RTD connection
4 terminal input
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Reference probe connection
6-pin Din with Infocon Technology
RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per°C
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C

Base Unit Specifications

Insert OD
24.4 mm (0.96 in)
Radial uniformity
±0.02°C at 50°C ±0.05°C at 420°C±0.10°C at 660°C
Display accuracy
±0.35°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C
Computer interface
RS-232 and 9930 Interface-it control software included
Stability
±0.03°C at 50°C ±0.04°C at 420°C±0.05°C at 660°C
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Immersion (well) depth
150 mm (5.9 in)
Power requirements
100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
Heating time
15 min: 50°C to 660°C
Display
LCD, °C or °F user-selectable
Cooling time
25 min: 660°C to 100°C 35 min: 660°C to 50°C
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Weight
7.7 kg (17 lb)
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Hysteresis
0.1
Temperature range at 23°C
50°C to 660°C (122°F to 1220°F)
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.015°C at 50°C ±0.025°C at 420°C ±0.035°C at 660°C
Resolution
0.01°
Axial uniformity at 40 mm (1.6 in)
±0.05°C at 50°C ±0.35°C at 420°C±0.5°C at 660°C