General specifications
- Size
- Height: 382 mm (15 in)Width: 242 mm (9.5 in)Depth: 400 mm (15.7 in)
- Weight
- 6109A: 16 kg (35 lb)7109A: 20 kg (44 lb)
- Fluid volume
- 2.5 liters (0.66 gallons)
- Fluid working area
- 75 mm x 75 mm (3 in x 3 in)
- Maximum fluid depth
- 154 mm (6.1 in)
- Factory calibration
- Traceable NVLAP-accredited calibration included
Specifications
- Range (at 25 °C ambient temperature)
- 35 °C to 250 °C: –25 °C to 140 °C (–15 °C cover off)
- Accuracy (maximum permissible error)
- 0.1 °C: 0.1 °C
- Display resolution 0.01 °, 0.001°
- 0.01 °, 0.001°
- Stability
- 0.015 °C: 0.01 °C
- Typical uniformity
- 0.03 °C up to 200 °C0.04 °C above 200 °C: 0.02 °C
- Repeatability
- 0.04 °C: 0.04 °C
- Typical heating time
- 35 °C to 100 °C: 25 minutes100 °C to 250 °C: 45 minutes: –25 °C to 25 °C: 35 minutes25 °C to 140 °C: 55 minutes
- Typical cooling time
- 250 °C to 100 °C: 35 minutes100 °C to 35 °C: 55 minutes: 140 °C to 25 °C: 45 minutes25 °C to –25 °C: 75 minutes
- Typical settling time
- 15 minutes: 10 minutes
Other Categories
ActuatorsBlocksCables, Pipes & HosesCompressorsControlCouplings & FittingsCylinders & PistonsElectronicsFiltersHVACIndustrial PC's & HMIsLightingLinear techniqueMotorsPower SupplyPower TransmissionPumps & ValvesRenewable energyRoboticsSeals, Rings & ConnectorsSensors & Measuring DevicesTools and AccessoriesCNCOther