Fluke

FLK-729JP 1M FC

1 / 3
thumbnail
thumbnail
thumbnail
Part Number
FLK-729JP 1M FC
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

Resolution

MA dc source, simulate and measure
1 μA
Voltage dc measurement
1 mV
Accuracy
0.01% ±2 LSD all ranges (at 23 °C ±5 °C)
Stability
20 ppm of full scale /°C from -10 °C to +18 °C and 28 °C to +50 °C
MA simulate external voltage requirement
12 V dc to 30 V dc
Loop compliance voltage
24 V dc @ 20 mA
Temperature measurement only/100 Ω Pt(385) RTD
-50 °C to 150 °C (-58 °F to 302 °F)
Temperature resolution
0.01 °C (0.01 °F)
Temperature accuracy
±0.1 °C (0.2 °F) ±0.25 °C (±0.45 °F) combined uncertainty when using 720 RTD probe (optional accessory)
Drive capability
1200 Ω without HART resistor, 950 Ω with internal HART resistor

Battery will only charge to 40 °C

Storage temperature
-40 °C to 60 °C
Operating altitude
3000 m
Storage altitude
13000 m

Wireless (729 FC only)

Radio frequency certification
FCC ID: T68-FBLE IC:6627A-FBLE
Wireless radio frequency range
2412 MHz to 2462 MHz

Model Specifications

KPa range, resolution
-82.73 to 1034.21 kPa
Comment
For Japan, wireless communication for Fluke Connect

Ranges

MA measure, source and simulate
0 mA to 24 mA
Volts dc measure
0 V dc to 30 V dc

Environmental specification

Operating temperature
-10 °C to 50 °C for measurement, 0 °C to 50 °C for pressure control
Operating temperature with battery
-10 °C to 40 °C

Mechanical specification

Size (HxWxL)
7 x 27.94 x 17.27 cm (2.75 x 11 x 6.8 in)
Weight
6.5 lbs., 2.95 kg
Ingress protection
IP54

Documents

Data Sheet

Download