Specifications
- Input channels
- 1
- Logging
- 25 readings with statistics
- Sample interval(normal)
- 1 second
- Sample interval (fast mode)
- 0.3 seconds (see technical manual for details)
- Sensor types
- PRTs, RTDs, Thermistors, and Thermocouples
- Thermocouple types
- C,E,J,K,L,M,N,T,U,B,R,S
- Operating temperature
- -10°C to 60°C (Best accuracy 13°C to 33°C)
- Power requirements
- 3 AA alkaline batteries
- Size
- 96 x 200 x 47 mm(3.75 x 7.9 x 1.86 in)
- Weight
- 0.65 kg (1.4 lb)
- Environmental conditions for best accuracy
- 13°C to 33°C
- Millivolt range and accuracy
- –10 mV to 75 mV ±(0.005% + 5 μV)
- Resistance range and accuracy
- 0 Ω to 400 Ω ±(0.004% + 0.002 Ω)200 Ω to 50 kΩ ±(0.01% + 0.5 Ω)50 kΩ to 500 kΩ ±(0.03%)
- Temperature coefficient, voltage (–10°C to 13°C , +33°C to 60°C)
- ±(0.001%/°C + 1 μV/°C)
- Temperature coefficient, resistance (–10°C to 13°C , +33°C to 60°C)
- 0.0008%/°C + 0.0004 Ω (0 Ω to 400 Ω)0.002%/°C + 0.1 Ω (0 Ω to 50 kΩ)0.06%/°C + 0.1 Ω (50 kΩ to 500 kΩ)
- Excitation current, resistance
- 1 mA (0 Ω to 400 Ω)10 μA (0 Ω to 50 kΩ)2 µA (50 kΩ to 500 kΩ)
- Type B
- ±0.85°C from 600°C to 800°C±0.68°C from 800°C to 1000°C±0.57°C from 1000°C to 1800°C
- Type C
- ±0.32°C from 100ºC to 550°C±0.71°C from 550°C to 2300°C
- Type E
- ±0.52°C from -200°C to 0°C±0.22°C from 0°C to 950°C
- Type J
- ±0.52°C from -200°C to 0°C±0.23°C from 0°C to 1200°C
- Type K
- ±0.61°C from -200°C to 0°C±0.24°C from 0°C to 1370°C
- Type L
- ±0.36°C from -200°C to 0°C±0.23°C from 0°C to 1370°C
- Type M
- ±0.26°C from -20°C to 0°C±0.25°C from 0°C to 400°C±0.22°C from 400°C to 1400°C
- Type N
- ±0.72°C from -200°C to 0°C±0.28°C from 0°C to 1300°C
- Type R
- ±1.09°C from -20°C to 0°C±0.97°C from 0°C to 500°C±0.49°C from 500°C to 1750°C
- Type S
- ±1.05°C from -20°C to 0°C±0.95°C from 0°C to 500°C±0.56°C from 500°C to 1750°C
- Type T
- ±0.60°C from -200°C to 0°C±0.25°C from 0°C to 400°C
- Type U
- ±0.54°C from -200°C to 0°C±0.24°C from 0°C to 400°C
Accuracies with Selected Probes (±°C)
- -200°C
- 5616-12: 0.0145615-6: 0.0255627A-12: 0.0275610-9: n/a
- 0°C
- ±0.002
- 100°C
- ±0.030
- 300°C
- 5616-12: 0.0405615-6: 0.0435627A-12: 0.1035610-9: n/a
- 420°C
- 5616-12: 0.0505615-6: n/a5627A-12: 0.1305610-9: n/a
- –100°C
- ±0.011
- 200°C
- ±0.023
- 400°C
- ±0.031
- 600°C
- ±0.039
- 25°C
- ±0.003
- 50°C
- ±0.006
- 75°C
- ±0.014
Documents
Data Sheet
DownloadOther Categories
ActuatorsBlocksCables, Pipes & HosesCompressorsControlCouplings & FittingsCylinders & PistonsElectronicsFiltersHVACIndustrial PC's & HMIsLightingLinear techniqueMotorsPower SupplyPower TransmissionPumps & ValvesRenewable energyRoboticsSeals, Rings & ConnectorsSensors & Measuring DevicesTools and AccessoriesCNCOther