Specifications

Range (at 23°C ambient)
–30°C to 155°C (–22°F to 311°F)
Display accuracy
±0.1°C full range
Stability
±0.005°C full range
Axial uniformity
±0.025°C at –30°C±0.02°C at 0°C ±0.07°C at 155°C
Radial uniformity
±0.01°C full range
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.005°C at –30°C±0.005°C at 0°C±0.01°C at 155°C
Hysteresis
0.025°C
Well depth
203 mm (8 in)
Resolution
0.001°C
Display
LCD, °C or °F, user-selectable
Key pad
Ten key with decimal and ± button. Function keys, menu key, and °C / °F key.
Cooling time
30 min: 23°C to –30°C25 min: 155°C to 23°C
Heating time
44 min: 23°C to 155°C56 min: –30°C to 155°C
Size (H x W x D)
366 x 203 x 323 mm (14.4 x 8 x 12.7 in)
Weight
15 kg (33 lb)
Power
115 V AC (±10%), or230 V AC (±10%), 50/60 Hz, 550 W
Computer interface
RS-232 Interface with 9930 Interface-it control software included
Traceable calibration (NIST)
Data at –30°C, 0°C, 50°C, 100°C, and 155°C
Specifications
Built-in Reference Input
Temperature range
–200°C to 962°C (–328°F to 1764°F)
Resistance range
0 Ω to 400 Ω, auto-ranging
Characterizations
ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 Callendar-Van Dusen (CVD): R0, a, b, d
Resistance accuracy
0 Ω to 20 Ω: 0.0005 W20 Ω to 400 Ω: 25 ppm
Temperature accuracy(does not include probe uncertainty)
10 Ω PRTs:±0.013°C at 0°C±0.014°C at 155°C±0.019°C at 425°C±0.028°C at 700°C: 25 Ω and 100 Ω PRTs:±0.005°C at –100°C±0.007°C at 0°C±0.011°C at 155°C±0.013°C at 225°C±0.019°C at 425°C±0.027°C at 661°C
Resistance resolution
0 Ω to 20 Ω: 0.0001 Ω20 Ω to 400 Ω: 0.001 Ω
Measurement period
1 second
Probe connection
4-wire with shield, 5-pin DIN connector
Calibration
NVLAP accredited (built-in reference input only), NIST-traceable calibration provided