Specifications
- Temperature range at 23°C
- –95°C to 140°C (–139°F to 284°F)
- Display accuracy
- ±0.2°C full range
- Accuracy with external reference3
- ±0.05°C full range
- Stability
- ±0.015°C full range
- Axial uniformity at 40 mm (1.6 in)
- ±0.05°C full range
- Radial gradient
- ±0.01°C full range
- Loading effect
- ±0.10°C at 140°C
- Operating conditions
- RH (non-condensing) < 2000 m altitude
- Environmental conditions for all specifications except temperature range
- 13°C to 33°C
- Immersion (well) depth
- 160 mm (6.3 in)
- Well diameter
- 30 mm (1.18 in)
- Heating time1
- –95°C to 140°C: 40 min
- Cooling time1
- 23°C to –90°C: 80 min
- Cooling time
- 140°C to 23°C: 60 min
- Stabilization time2
- 15 min
- Resolution
- 0.01°
- Display
- LCD, °C or °F user selectable
- Size (H x W x D)
- 480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in)
- Weight
- 16 kg (35 lb)
- Power requirements
- 200 V to 230 V (±10%) 50/60 Hz, 575 W
- System fuse ratings
- 230 V: 3.15 A T 250 V
- 4–20 mA fuse (-P model only)
- 50 mA F 250 V
- Computer interface
- RS-232, USB Serial, and 9930 interface-it temperature calibration software included
- Safety
- IEC 61010-1, Installation Category II, Pollution degree 2
- Electromagnetic environment
- IEC 61326-1: Basic
- Refrigerants
- < 20 g, ASHRAE safety group A1
P Specifications
- Built-in reference thermometer readout accuracy (4-wire reference probe)3
- ±0.010°C at –95°C
- Built-in reference thermometer readout accuracy (4-wire reference probe)
- ±0.025°C at 140°C
- Reference resistance range
- 0 Ω to 400 Ω
- Reference resistance accuracy4
- 0 Ω to 42 Ω: ±0.0025 Ω42 Ω to 400 Ω: ±60 ppm of reading
- Reference characterizations
- ITS-90, CVD, IEC-751, resistance
- Reference measurement capability
- 4 wire
- Reference probe connection
- 6-pin din with INFO-CON technology
- Built-in RTD thermometer readout accuracy
- PT-100 (JIS): ±0.02°C at 0°C
- RTD resistance range
- 0 Ω to 400 Ω
- Resistance accuracy4
- 0 Ω to 25 Ω: ±0.002 Ω
- Resistance accuracy
- 25 Ω to 400 Ω: ±80 ppm of reading
- RTD characterizations
- PT-100 (385), (JIS), (3926), NI-120, resistance
- RTD measurement capability
- 2-wire, 3-wire, and 4-wire RTD with jumpers only
- RTD connection
- 4-terminal input
- Built-in TC thermometer readout accuracy5
- Type J: ±0.70°C at 140°C
- Built-in TC thermometer readout accuracy
- Type C: ±1.00°C at 140°C
- TC millivolt range
- –10 mV to 75 mV
- Voltage accuracy
- 0.025% of reading +0.01 mV
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- TC connection
- Miniature connectors (ASTM E1684)
- Built-in mA readout accuracy
- 0.02% of reading + 0.002 mA
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- MA connection
- 2 terminal input
- Loop power function
- 24 V DC loop power
- Built-in electronics temperature coefficient(0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per °C
Other Categories
ActuatorsBlocksCables, Pipes & HosesCompressorsControlCouplings & FittingsCylinders & PistonsElectronicsFiltersHVACIndustrial PC's & HMIsLightingLinear techniqueMotorsPower SupplyPower TransmissionPumps & ValvesRenewable energyRoboticsSeals, Rings & ConnectorsSensors & Measuring DevicesTools and AccessoriesCNCOther