P Specifications

Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.010°C at –95°C
Built-in reference thermometer readout accuracy (4-wire reference probe)
±0.025°C at 140°C
Reference resistance range
0 Ω to 400 Ω
Reference resistance accuracy4
0 Ω to 42 Ω: ±0.0025 Ω42 Ω to 400 Ω: ±60 ppm of reading
Reference characterizations
ITS-90, CVD, IEC-751, resistance
Reference measurement capability
4 wire
Reference probe connection
6-pin din with INFO-CON technology
Built-in RTD thermometer readout accuracy
PT-100 (JIS): ±0.02°C at 0°C
RTD resistance range
0 Ω to 400 Ω
Resistance accuracy4
0 Ω to 25 Ω: ±0.002 Ω
Resistance accuracy
25 Ω to 400 Ω: ±80 ppm of reading
RTD characterizations
PT-100 (385), (JIS), (3926), NI-120, resistance
RTD measurement capability
2-wire, 3-wire, and 4-wire RTD with jumpers only
RTD connection
4-terminal input
Built-in TC thermometer readout accuracy5
Type J: ±0.70°C at 140°C
Built-in TC thermometer readout accuracy
Type C: ±1.00°C at 140°C
TC millivolt range
–10 mV to 75 mV
Voltage accuracy
0.025% of reading +0.01 mV
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
TC connection
Miniature connectors (ASTM E1684)
Built-in mA readout accuracy
0.02% of reading + 0.002 mA
MA range
Cal 4-22 mA, Spec 4-24 mA
MA connection
2 terminal input
Loop power function
24 V DC loop power
Built-in electronics temperature coefficient(0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C

Specifications

Temperature range at 23°C
–95°C to 140°C (–139°F to 284°F)
Display accuracy
±0.2°C full range
Accuracy with external reference3
±0.05°C full range
Stability
±0.015°C full range
Axial uniformity at 40 mm (1.6 in)
±0.05°C full range
Radial gradient
±0.01°C full range
Loading effect
±0.10°C at 140°C
Operating conditions
RH (non-condensing) < 2000 m altitude
Environmental conditions for all specifications except temperature range
13°C to 33°C
Immersion (well) depth
160 mm (6.3 in)
Well diameter
30 mm (1.18 in)
Heating time1
–95°C to 140°C: 40 min
Cooling time1
23°C to –90°C: 80 min
Cooling time
140°C to 23°C: 60 min
Stabilization time2
15 min
Resolution
0.01°
Display
LCD, °C or °F user selectable
Size (H x W x D)
480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in)
Weight
16 kg (35 lb)
Power requirements
200 V to 230 V (±10%) 50/60 Hz, 575 W
System fuse ratings
230 V: 3.15 A T 250 V
4–20 mA fuse (-P model only)
50 mA F 250 V
Computer interface
RS-232, USB Serial, and 9930 interface-it temperature calibration software included
Safety
IEC 61010-1, Installation Category II, Pollution degree 2
Electromagnetic environment
IEC 61326-1: Basic
Refrigerants
< 20 g, ASHRAE safety group A1