Fluke

P3125-4

1 / 1
thumbnail
Part Number
P3125-4
  • Get up to 90 days payment terms from 500+ suppliers
  • Reselling discounts up to 40%!
Sign-Up
Send request for price
1
Payment methods:
Request a Quote and we will send you information on availability and price.

Operating Fluid

Mineral oil
Shell spindle oil 22, our reference 55-655, viscosity 40 cs @ 20°C

Pressure Ranges

Oil operated
Ranges to 20,000 psi (1400 bar)

Accuracy

Accuracy1
±0.015% of reading (±0.008% optional)
Accuracy1
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range.

Weight Increments

Optional fractional weights
Up to 500 psi (35 bar): 0.1 psi (0.01 bar) 500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar) 5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar)
Minimum standard weight increments
Up to 500 psi (35 bar): 1 psi (0.1 bar) 500 to 5000 psi (35 to 350 bar): 10 psi (1 bar) 5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar) Model P3112: 1 psi (0.1 bar)

Materials of Construction

Weight density
2.7 g/cm³
Piston material
Tungsten carbide with nickel binder Density- 15.0 g/cm³
Standard weight material
Series 3 non-magnetic austenitic stainless steel
Optional fractional weights
Solution heat treated aluminum
Cylinder material
Hardened martensitic steel Tungsten carbide (water above 500 psi, 35 bar)
Thermal coefficients of expansion
Oil piston/cylinder 16.5 ppm/°C Water (above 500 psi, 35 bar) 11 ppm/°C
Weight density
7.8 g/cm³

Options

Fluids
Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory.
PressCal software
Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments.

General

Instrument size (W x D x H)
17.5 x 12 x 8.5
Reservoir volume
150 cc
Screw press displacement
5.5 cc
Pump displacement
4.7 cc per stroke
Test port adaptors
1/8, 1/4, 3/8 and ½ NPT or BSP
Instrument weight
36 lbs (16 kg)
Mass set weight (typical)
80 lbs (36 kg)
O-ring seal materials
Buna N as standard, Viton and EPDM available