Fluke
9190A-F
1 / 4
- Part Number
- 9190A-F
Payment methods:
- Best offer from:
- Trusted Automa.Net Supplier
- Availability:
- 1 pc.In-StockUpdated at: 17-09-2024
- Warranty:
- 12 months
- Estimated Delivery:
- On Request
- Condition:
- New Factory Sealed
New Factory Sealed - A brand-new, unused, unopened, undamaged item in its original manufacturer-sealed packaging.
Secure Payment Process
Return & Refund Policies
Shipping Policies
Warranty Policies
Specifications
- Loading effect
- ±0.10°C at 140°C
- Cooling time1
- 23°C to –95°C: 90 min
- Power requirements
- 200 V to 230 V (±10%) 50/60 Hz, 575 W
- System fuse ratings
- 230 V: 3.15 A T 250 V
- Weight
- 16 kg (35 lb)
- Refrigerants
- < 20 g, ASHRAE safety group A2L
- Cooling time1
- 140°C to 23°C: 60 min
- Loading effect
- (versus display with one 6.35 mm probe)
- Well diameter
- 30 mm (1.18 in)
- Electromagnetic environment
- IEC 61326-1: Basic
- Cooling time1
- 23°C to –90°C: 80 min
- Stability
- ±0.015°C full range
- Radial gradient
- ±0.01°C full range
- Refrigerants
- R32 (Difluoromethane)
- Environmental conditions for all specifications except temperature range
- 13°C to 33°C
- Stabilization time2
- 15 min
- Immersion (well) depth
- 160 mm (6.3 in)
- Size (H x W x D)
- 480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in)
- Axial uniformity at 40 mm (1.6 in)
- ±0.05°C full range
- 4–20 mA fuse (-P model only)
- 50 mA F 250 V
- Refrigerants
- R704 (Helium)
- Computer interface
- RS-232, USB Serial, and 9930 interface-it temperature calibration software included
- Safety
- IEC 61010-1, Installation Category II, Pollution degree 2
- Loading effect
- ±0.25°C at –95°C
- System fuse ratings
- 115 V: 6.3 A T 250 V
- Resolution
- 0.01°
- Refrigerants
- < 20 g, ASHRAE safety group A1
- Temperature range at 23°C
- –95°C to 140°C (–139°F to 284°F)
- Power requirements
- 100 V to 115 V (±10%) 50/60 Hz, 575 W
- Display
- LCD, °C or °F user selectable
- Operating conditions
- 0°C to 35°C, 0% to 90%
- Display accuracy
- ±0.2°C full range
- Accuracy with external reference3
- ±0.05°C full range
- Loading effect
- (with a 6.35 mm reference probe and three 6.35 mm probes)
- Operating conditions
- RH (non-condensing) < 2000 m altitude
- Loading effect
- ±0.006°C full range
- Heating time1
- –95°C to 140°C: 40 min
P Specifications
- MA connection
- 2 terminal input
- Reference resistance range
- 0 Ω to 400 Ω
- Built-in RTD thermometer readout accuracy
- PT-100 (3926): ±0.02°C at 0°C
- Built-in RTD thermometer readout accuracy
- PT-100 (385): ±0.02°C at 0°C
- Built-in TC thermometer readout accuracy5
- Type S: ±1.60°C at 140°C
- RTD characterizations
- PT-100 (385), (JIS), (3926), NI-120, resistance
- Built-in TC thermometer readout accuracy5
- Type C: ±1.00°C at 140°C
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- 1. For ambient temperature of 23°C. 2. Time from when the SETPOINT is reached to when the unit is with in stability specification. 3. The temperature range may be limited by the reference probe connected to the readout. The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 4. Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05 Ω to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires. 5. The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz.
- Built-in TC thermometer readout accuracy5
- Type T: ±0.60°C at 140°C
- Built-in mA readout accuracy
- 0.02% of reading + 0.002 mA
- Built-in reference thermometer readout accuracy (4-wire reference probe)3
- ±0.013°C at –25°C
- Built-in TC thermometer readout accuracy5
- Type U: ±0.70°C at 140°C
- Internal cold junction compensation accuracy
- ±0.35°C (ambient of 13°C to 33°C)
- TC connection
- Miniature connectors (ASTM E1684)
- Built-in TC thermometer readout accuracy5
- Type M: ±0.65°C at 140°C
- Built-in TC thermometer readout accuracy5
- Type E: ±0.60°C at 140°C
- TC millivolt range
- –10 mV to 75 mV
- Built-in TC thermometer readout accuracy5
- Type L: ±0.65°C at 140°C
- Voltage accuracy
- 0.025% of reading +0.01 mV
- Built-in TC thermometer readout accuracy5
- Type K: ±0.75°C at 140°C
- Built-in RTD thermometer readout accuracy
- PT-100 (JIS): ±0.02°C at 0°C
- Built-in TC thermometer readout accuracy5
- Type J: ±0.70°C at 140°C
- Reference resistance accuracy4
- 0 Ω to 42 Ω: ±0.0025 Ω 42 Ω to 400 Ω: ±60 ppm of reading
- Built-in reference thermometer readout accuracy (4-wire reference probe)3
- ±0.020°C at 50°C
- Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
- ±0.005% of range per °C
- Built-in TC thermometer readout accuracy5
- Type R: ±1.60°C at 140°C
- RTD resistance range
- 0 Ω to 400 Ω
- RTD connection
- 4-terminal input
- Reference probe connection
- 6-pin din with INFO-CON technology
- Resistance accuracy4
- 25 Ω to 400 Ω: ±80 ppm of reading
- Reference characterizations
- ITS-90, CVD, IEC-751, resistance
- Resistance accuracy4
- 0 Ω to 25 Ω: ±0.002 Ω
- Built-in reference thermometer readout accuracy (4-wire reference probe)3
- ±0.015°C at 0°C
- MA range
- Cal 4-22 mA, Spec 4-24 mA
- Built-in reference thermometer readout accuracy (4-wire reference probe)3
- ±0.025°C at 140°C
- Reference measurement capability
- 4 wire
- Built-in RTD thermometer readout accuracy
- NI-120: ±0.015°C at 0°C
- RTD measurement capability
- 2-wire, 3-wire, and 4-wire RTD with jumpers only
- Built-in TC thermometer readout accuracy5
- Type N: ±0.75°C at 140°C
- Built-in reference thermometer readout accuracy (4-wire reference probe)3
- ±0.010°C at –95°C
- Loop power function
- 24 V DC loop power
Other Categories
ActuatorsBlocksCables, Pipes & HosesCompressorsControlCouplings & FittingsCylinders & PistonsElectronicsFiltersHVACIndustrial PC's & HMIsLightingLinear techniqueMotorsPower SupplyPower TransmissionPumps & ValvesRenewable energyRoboticsSeals, Rings & ConnectorsSensors & Measuring DevicesTools and AccessoriesCNCOther