Fluke

9190A-F

1 / 4
thumbnail
thumbnail
thumbnail
thumbnail
Part Number
9190A-F
Payment methods:
Best offer from:
PLTrusted Automa.Net Supplier
Availability:
1 pc.
In-Stock
Updated at: 17-09-2024
Warranty:
12 months
Estimated Delivery:
On Request
Condition:
New Factory Sealed
New Factory Sealed - A brand-new, unused, unopened, undamaged item in its original manufacturer-sealed packaging.
Secure Payment Process
Return & Refund Policies
Shipping Policies
Warranty Policies

Specifications

Loading effect
±0.10°C at 140°C
Cooling time1
23°C to –95°C: 90 min
Power requirements
200 V to 230 V (±10%) 50/60 Hz, 575 W
System fuse ratings
230 V: 3.15 A T 250 V
Weight
16 kg (35 lb)
Refrigerants
< 20 g, ASHRAE safety group A2L
Cooling time1
140°C to 23°C: 60 min
Loading effect
(versus display with one 6.35 mm probe)
Well diameter
30 mm (1.18 in)
Electromagnetic environment
IEC 61326-1: Basic
Cooling time1
23°C to –90°C: 80 min
Stability
±0.015°C full range
Radial gradient
±0.01°C full range
Refrigerants
R32 (Difluoromethane)
Environmental conditions for all specifications except temperature range
13°C to 33°C
Stabilization time2
15 min
Immersion (well) depth
160 mm (6.3 in)
Size (H x W x D)
480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in)
Axial uniformity at 40 mm (1.6 in)
±0.05°C full range
4–20 mA fuse (-P model only)
50 mA F 250 V
Refrigerants
R704 (Helium)
Computer interface
RS-232, USB Serial, and 9930 interface-it temperature calibration software included
Safety
IEC 61010-1, Installation Category II, Pollution degree 2
Loading effect
±0.25°C at –95°C
System fuse ratings
115 V: 6.3 A T 250 V
Resolution
0.01°
Refrigerants
< 20 g, ASHRAE safety group A1
Temperature range at 23°C
–95°C to 140°C (–139°F to 284°F)
Power requirements
100 V to 115 V (±10%) 50/60 Hz, 575 W
Display
LCD, °C or °F user selectable
Operating conditions
0°C to 35°C, 0% to 90%
Display accuracy
±0.2°C full range
Accuracy with external reference3
±0.05°C full range
Loading effect
(with a 6.35 mm reference probe and three 6.35 mm probes)
Operating conditions
RH (non-condensing) < 2000 m altitude
Loading effect
±0.006°C full range
Heating time1
–95°C to 140°C: 40 min

P Specifications

MA connection
2 terminal input
Reference resistance range
0 Ω to 400 Ω
Built-in RTD thermometer readout accuracy
PT-100 (3926): ±0.02°C at 0°C
Built-in RTD thermometer readout accuracy
PT-100 (385): ±0.02°C at 0°C
Built-in TC thermometer readout accuracy5
Type S: ±1.60°C at 140°C
RTD characterizations
PT-100 (385), (JIS), (3926), NI-120, resistance
Built-in TC thermometer readout accuracy5
Type C: ±1.00°C at 140°C
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
1. For ambient temperature of 23°C. 2. Time from when the SETPOINT is reached to when the unit is with in stability specification. 3. The temperature range may be limited by the reference probe connected to the readout. The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 4. Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05 Ω to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires. 5. The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz.
Built-in TC thermometer readout accuracy5
Type T: ±0.60°C at 140°C
Built-in mA readout accuracy
0.02% of reading + 0.002 mA
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.013°C at –25°C
Built-in TC thermometer readout accuracy5
Type U: ±0.70°C at 140°C
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
TC connection
Miniature connectors (ASTM E1684)
Built-in TC thermometer readout accuracy5
Type M: ±0.65°C at 140°C
Built-in TC thermometer readout accuracy5
Type E: ±0.60°C at 140°C
TC millivolt range
–10 mV to 75 mV
Built-in TC thermometer readout accuracy5
Type L: ±0.65°C at 140°C
Voltage accuracy
0.025% of reading +0.01 mV
Built-in TC thermometer readout accuracy5
Type K: ±0.75°C at 140°C
Built-in RTD thermometer readout accuracy
PT-100 (JIS): ±0.02°C at 0°C
Built-in TC thermometer readout accuracy5
Type J: ±0.70°C at 140°C
Reference resistance accuracy4
0 Ω to 42 Ω: ±0.0025 Ω 42 Ω to 400 Ω: ±60 ppm of reading
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.020°C at 50°C
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C
Built-in TC thermometer readout accuracy5
Type R: ±1.60°C at 140°C
RTD resistance range
0 Ω to 400 Ω
RTD connection
4-terminal input
Reference probe connection
6-pin din with INFO-CON technology
Resistance accuracy4
25 Ω to 400 Ω: ±80 ppm of reading
Reference characterizations
ITS-90, CVD, IEC-751, resistance
Resistance accuracy4
0 Ω to 25 Ω: ±0.002 Ω
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.015°C at 0°C
MA range
Cal 4-22 mA, Spec 4-24 mA
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.025°C at 140°C
Reference measurement capability
4 wire
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C
RTD measurement capability
2-wire, 3-wire, and 4-wire RTD with jumpers only
Built-in TC thermometer readout accuracy5
Type N: ±0.75°C at 140°C
Built-in reference thermometer readout accuracy (4-wire reference probe)3
±0.010°C at –95°C
Loop power function
24 V DC loop power